Previous Article | Next Article ![]()
Clinical and Diagnostic Laboratory Immunology, January 2004, p. 50-55, Vol. 11, No. 1
1071-412X/04/$08.00+0 DOI: 10.1128/CDLI.11.1.50-55.2004
Biological Monitoring Laboratory Section, Biomonitoring and Health Assessment Branch, National Institute for Occupational Safety and Health, Cincinnati, Ohio,1 Microbial Pathogenesis and Immune Response Laboratory, National Center for Infectious Diseases, Centers for Disease Control and Prevention, Atlanta, Georgia2
Received 4 August 2003/ Returned for modification 12 September 2003/ Accepted 2 October 2003
|
|
|---|
|
|
|---|
|
|
|---|
Antigens. For the ELISA, recombinant anthrax toxin PA with an amino acid sequence concurring with that from the B. anthracis V770-NP1-R anthrax vaccine strain was obtained from the National Institute of Craniofacial and Dental Research, National Institutes of Health, Bethesda, Md. Antigen was produced and purified as described previously (9, 12) and was stored frozen at -80°C in small aliquots (10 to 100 µl, 4.75 mg/ml) in 5 mM HEPES (pH 7.3).
The antigens used in the FCMIA were recombinant PA and LF obtained from List Biological Laboratories, Inc., (Campbell, Calif.). Both antigens migrated as single major bands with apparent molecular masses of 83,000 Da (PA) and 90,000 Da (LF) on 10% polyacrylamide gels in the presence of sodium dodecyl sulfate. The PA and LF were reconstituted in distilled water and stored as aliquots at -20°C. Before use, individual aliquots were thawed, mixed thoroughly, and used immediately.
ELISA procedure. The ELISA procedure has been described in detail previously (15). Serum standards and sera for testing were prepared at the appropriate dilutions in phosphate-buffered saline (PBS) containing 5% skim milk and 0.5% Tween 20 (pH 7.4). The human standard reference serum and test sera were serially diluted twofold in the same buffer solution in the plate. The minimum dilution of test serum was 1:50. The final volume in all wells was 100 µl. Bound anti-PA IgG was detected by using horseradish peroxidase-conjugated mouse anti-human IgG Fc PAN monoclonal HP6043 and 2,2'-azino-di(3-ethyl-benzthiazoline-6-sulfonate) (ABTS)-H2O2 substrate (100 µl/well). Color development was carried out over 30 min (±5 min) and was stopped by addition of 100 µl of peroxidase stop solution (Kirkegaard & Perry Laboratories, Gaithersburg, Md.) to all wells of the test plates. Optical densities values were read within 30 min of addition of the stop solution with an MRX Revelation microtiter plate reader (Thermo Labsystems, Franklin, Mass.) at a wavelength of 410 nm with a 610-nm reference filter. Data were analyzed by using a four-parameter (4-PL) logistic-log curve fitting model with ELISA for Windows software (15). A calibration factor for the standard reference serum was used to determine the concentration of anti-PA IgG in micrograms per milliliter of serum. Comparison sera were diluted 1:50.
Preparation of PA and LF coupled microspheres. Briefly, after being washed twice with 80 µl of activation buffer (0.1M NaH2PO4 [pH 6.2]), two sets of spectrally differentiable carboxylated microspheres (2.5 x 106; Luminex Corp., Austin, Tex.) were pelleted (5,000 x g for 2 min) in 1.5-ml centrifuge tubes in a microcentrifuge (Eppendorf, Hamburg, Germany). The microspheres were resuspended by sonication (mini sonicator; Cole Parmer, Vernon Hills, Ill.) and gentle vortexing (VWR International, West Chester, Pa.) in 80 µl of activation buffer to which 10 µl of activation buffer containing 50 mg of 1-ethyl-3-(3-dimethylaminopropyl)carbodiimide hydrochloride (EDC; Pierce Chemical Co., Rockford, Ill.) per ml and 10 µl of activation buffer containing 50 mg of N-hydroxysulfosuccinimide, sodium salt (sulfo-NHS; Pierce Chemical Co.) per ml were added. The mixture was allowed to incubate for 20 min at room temperature. The microspheres were then washed twice in 500 µl of coupling buffer [0.05 M 2-(N-morpholino)ethanesulfonic acid (MES) (Sigma Chemical Co.); pH 5.0], a solution of PA or LF (40 µg/ml) in 500 µl of coupling buffer was added, and the mixture was incubated for 2 h at room temperature. The 40-µg/ml concentration was used for coupling because preliminary experiments indicated that this coupling concentration yielded high mean fluorescence intensities (MFIs) and good inhibition (data not shown). The coupled microspheres were then washed twice in 1 ml of wash buffer (PBS containing 0.05% Tween 20 [Sigma no. 3563]) and stored in 0.5 ml of storage buffer (PBS, 1% bovine serum albumin [Sigma no. 3688], 0.05% sodium azide [EM Sciences, Cherry Hill, N.J.] [pH 7.4]). Microsphere concentrations were determined using a hemacytometer (Bright Line; VWR International).
Monoplex and multiplexed measurements obtained using FCMIA. For comparison with the anti-PA IgG ELISA, a monoplex anti-PA IgG FCMIA format was used with pooled sera obtained from anthrax vaccinees (AVR414). Serum AVR414, however, contains only low concentrations of anti-LF IgG, making it unsuitable for use as a standard reagent for determining the concentrations of that antibody. For this reason, serum AVR733, which was obtained from a patient with a confirmed case of human anthrax and which contains both anti-PA and anti-LF IgG, was used in a multiplexed FCMIA format employing a duplex mixture of microspheres. Briefly, suspensions of microspheres (100 microspheres/µl) were incubated for 1 h at 37°C in blocking buffer (PBS [Sigma no. 3813], 5% dried skim milk [Difco. Sparks, Md.], 0.05% sodium azide [pH 7.4]) to minimize non-specific binding during subsequent steps. An aliquot (50 µl) of this suspension (approximately 5,000 microspheres) was added to the wells of a 1.2-µm-pore-size filter membrane microtiter plate (no. MABVN1250; Millipore Corp., Bedford, Mass.), and the liquid was aspirated by use of a vacuum manifold filtration system (Millipore no. MAVM09601). The microspheres were then washed three times with 200 µl of wash buffer, and each wash followed by vacuum aspiration. To run an experiment, 50-µl portions (in duplicate) of diluted comparison sera (1:250 in dilution buffer consisting of PBS, 5% dried skim milk, 0.05% sodium azide, and 1% Tween 20 [pH 7.4]), diluted standards prepared from AVR 414, dilutions of serum AVR733, or reagent blank were added to the microspheres in the wells of the filter membrane microtiter plate and incubated for 30 min at 37°C with shaking. The liquid was vacuum aspirated, and the wells were again washed three times with 200 µl of wash buffer. An aliquot (50 µl) of a 2-µg/ml solution of detection antibody (mouse anti-human IgG-red algae-phycoerythrin, clone HP6043 [Biotrend Chemicals, Inc., Destin, Fla.]) in dilution buffer was added to the wells of the plate and incubated for 15 min at 37°C. The wells were again washed three times with 200 µl of wash buffer and resuspended in 100 µl of wash buffer. The plate was shaken vigorously for approximately 1 min to disperse the microspheres and was placed into the autosampler platform of the Luminex 100 instrument (Luminex 100 flow analyzer, coupled with a 96-well plate autosampler, Luminex XYP [Luminex]) as previously described (3), using software, calibration microspheres, and sheath fluid supplied by the manufacturer. The instrument was programmed to collect data from 100 microspheres (classified by their internal fluorescence ratio) and acquire the MFI of the microsphere-PA-IgG(-LF)-anti-IgG-R-phycoerythrin complex.
Competitive inhibition FCMIA.
To determine the specificity of measurements performed by FCMIA, all sera which were used for the comparison with ELISA were preincubated with PA. For the multiplex analyses, serum AVR733 was preincubated with PA and LF (competitive inhibition). To perform an inhibition experiment, sera (final volume, 130 µl; final dilution, 1:250) were treated with either 80 µg of PA (AVR414) or PA and LF per ml singly or in combination (80 µg/ml each) or treated with dilution buffer and incubated overnight at 4°C. The sera were then centrifuged, and the supernatants were analyzed as outlined above. Inhibition was calculated by the following expression:
![]() |
Data analysis.
Standard curves were constructed from 4-PL logistic log fits of MFI versus the concentration (in micrograms per milliliter) of anti-PA IgG (AVR414) or anti-PA IgG and anti-LF (AVR733) IgG (11, 14) (SigmaPlot; SPSS, Chicago, Ill.). Duplicate results from individual samples were averaged, and concentrations were interpolated from the standard curve. The FCMIA was run three separate times, and the ELISA was run two or three separate times. Assessment of the "goodness of fit" and the dynamic range of the standards data to the 4-PL model was performed by standards recovery (10), evaluated by diluting AVR414 or AVR733 (in dilution buffer) from 1:25 to 1:1,000,000 (AVR414) or 1:25 to 1:100,000 (AVR733) and determining the resultant plot of measured concentration versus expected concentration by regression analyses of interpolated results from the 4-PL fit for linearity by regression. Recovery was calculated by the following equation:
![]() |
0.05 were considered statistically significant. |
|
|---|
![]() View larger version (21K): [in a new window] |
FIG. 1. (A) Representative FCMIA 4-PL (solid line) of the anti-PA IgG concentration versus MFI. The dotted lines represent the upper and lower 95% CI of the regression. The regression coefficient was r2 = 0.999 (P < 0.001). Data points are the mean of duplicate determinations. (B) Magnified view of panel A, showing a graphic representation of the MDC and RDL. The MDC is the concentration of anti-PA IgG corresponding to the interpolated intersection of the lower asymptote of the upper 95% confidence limit with the 4-PL logistic log fit of the standard curve data. The RDL is the concentration of anti-PA antibody corresponding to the interpolated intersection of the lower asymptote of the upper 95% confidence limit with the lower 95% confidence limit of the standard's data.
|
![]() View larger version (16K): [in a new window] |
FIG. 2. Linear regression of the results of diluting standard sera AVR414 from 1:50 to 1:100,000 and evaluating the resultant plot of measured concentration versus expected concentration for linearity. Data points are the mean of duplicate determinations.
|
![]() View larger version (16K): [in a new window] |
FIG. 3. Linear regression of anti-PA IgG concentrations measured by FCMIA and ELISA from 20 double-masked sera samples. Data points are the mean of two or three determinations performed in duplicate.
|
![]() View larger version (16K): [in a new window] |
FIG. 4. Representative multiplexed FCMIA 4-PL (solid line) of the concentration of anti-PA IgG and anti-LF IgG versus MFI. The goodness of fit, as described by the regression coefficients, was r2 = 0.999 (P < 0.0001) for both curves. Data points are the mean of duplicate determinations.
|
![]() View larger version (17K): [in a new window] |
FIG. 5. Linear regression of the results of diluting clinical anthrax serum AVR733 from 1:25 to 1:100,000 and evaluating the resultant plot of measured concentration versus expected concentration for LF (A) and PA (B). Data points are the mean of duplicate determinations.
|
|
|
|---|
The anti-PA FCMIA reported in the present study has intra- and interassay coefficients of variation of 5.7 and 13.1%, respectively, well within the <10 and <20% intra- and inter-assay guidelines for this type of assay (7), suggesting extremely good precision (15). The FCMIA also yielded a highly significant linear relationship on dilution and yielded a mean recovery of 101.7% ± 5.3% (SD) over the concentration range of 0.002 to 6.8 µg of anti-PA IgG per ml, suggesting that an accurate measurement of anti-PA IgG with an excellent dynamic range was performed by this method. The method MDC was 0.006 µg/ml, the RDL was 0.020 µg/ml, and the whole-serum equivalent MDC was 1.5 µg/ml, indicating greater sensitivity than that of ELISA. Investigation of the stability of the PA-coupled microspheres after storage yielded no significant differences in 4-PL anti-PA IgG standard curves (P > 0.10) after storage at 4°C for 9 months. Reagent stability for extended periods is desirable to be able to quickly respond to urgent requests for diagnostic testing. Preadsorption of the serum samples with PA yielded a mean inhibition of measured anti-PA IgG concentrations of 97.9% ± 1.9%, indicating immunochemical specificity. Anti-PA IgG concentrations calculated for the standard reagent set by FCMIA had a high positive correlation with those generated by ELISA (r2 = 0.852; P < 0.0001), and the interassay CV was 15.7% when the results from both assays for the masked sera samples (N = 20) were compared. Finally, analyses of the concentrations of anti-PA IgG found for identical samples measured by ELISA and FCMIA gave a linear regression line which was indistinguishable from a line of identity, indicating the equivalence of the methods. These data strongly suggest that FCMIA and ELISA are equivalent for the measurement of anti-PA IgG. They also indicate that FCMIA for anti-PA IgG, similar to ELISA, is accurate, reproducible, sensitive, stable, and precise. Discrepancies between FCMIA and ELISA became apparent at the lower limits of analyte detection (MDC and RDL). One sample, with an anti-PA IgG concentration by FCMIA of 2.4 µg/ml was above the MDC of the FCMIA (1.5 µg/ml) but below both the MDC and RDL of ELISA (3.0 and 4.5 µg/ml, respectively). This sample yielded an inhibition of 5.1% on preadsorption, which did not meet previously described criteria (15) for a "true"-positive sample; as such, it would be considered a false-positive result in FCMIA.
An inherent benefit of FCMIA over ELISA is its ability to be multiplexed (i.e., to measure numerous analytes simultaneously) (2, 4, 13). We demonstrated this by measuring anti-PA and anti-LF IgG levels simultaneously in a serum sample from a patient with a confirmed clinical anthrax infection. Both PA and LF, when measured as a multiplex, yielded highly linear responses on dilution and high mean percent recoveries (100.4 and 100.7%, respectively) and showed good inhibition from preadsorption with PA and LF (>85%). Anti-PA, -LF, and -endema factor IgG have been measured in sera from individuals with cutaneous and oral or oropharyngeal anthrax by three separate ELISAs (16); however, we believe this to be the first demonstration of a multiplexed measurement of anti-PA and anti-LF IgG.
According to the U.S. State Department and Department of Defense, the future deployment of biological agents by foreign militaries or terrorists is likely (19). Multiplexing becomes important in the postbioterrorism incident scenario, where multiple chemical and/or biological agents may be used simultaneously or sequentially. This scenario could potentially result in a need for numerous fast, accurate, and precise biological markers of infection by or exposure to many biological and chemical agents. Multiplexed FCMIA technology is unique in that both chemical and biological assays potentially could be performed simultaneously in one assay with the Luminex xMap technology reported on in the present work, an aspect we are presently investigating by using assays to detect immune responses from infection by or exposure to Francisella tularensis, Yersinia pestis, ricin, botulism toxin, staphylococcal enterotoxin B, and B. anthracis. Existing technologies (ELISA, gas-liquid chromatography/mass spectrometry, etc.) would have to be performed sequentially or simultaneously to measure numerous analytes. As the number of potentially exposed individuals and the number of potential analytes increases, use of these existing technologies could present significant resource and labor limitations. At present, numerous public health laboratories possess the necessary equipment and knowledge to perform ELISA testing for exposure to or infection by biological agents of terror. Incorporating multiplexed FCMIA technology into public health laboratories would involve the acquisition of specialized equipment and additional specific training; however, it may improve their capacity to respond to future, possibly more complex bioterrorism events.
In conclusion, we have described a FCMIA for anti-PA IgG in human serum which is accurate, precise, sensitive, and specific and agrees well with a specific, sensitive, and quantitative ELISA (15) when sera were analyzed in a double-masked analyses by the two methods. Compared with ELISA, FMCIA had improved sensitivity (ELISA MDC = 3.0 µg/ml; FCMIA MDC = 1.5 µg/ml) and a greater dynamic range (ELISA, 0.06 to 1.75 µg/ml; FCMIA, 0.006 to 6.8 µg/ml). The FCMIA is also faster, uses less sample, and has reagents which are stable for up to 9 months at 4°C. It has the ability to be multiplexed, as evidenced by the demonstration of the simultaneous measurement of anti-PA and anti-LF IgG in serum from patient with a confirmed anthrax case. The multiplexing capability of FCMIA is presently being exploited to detect immune responses from exposure to F. tularensis, Y. pestis, ricin, botulism toxin, and staphylococcal enterotoxin B as well as to measure antibody responses to B. anthracis.
Mention of a product or company name does not constitute endorsement by CDC.
We thank Ed Kreig (NIOSH) for assistance in performing some of the statistical analyses and Brian D. Plikaytis and Thomas H. Taylor, Jr., of DBMD, NCID, for critical review of the manuscript.
|
|
|---|
This article has been cited by other articles:
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Copyright © 2009 by the American Society for Microbiology. For an alternate route to Journals.ASM.org, visit: http://intl-journals.asm.org | More Info»