Clinical and Diagnostic Laboratory Immunology, March 2004, p. 255-260, Vol. 11, No. 2
1071-412X/04/$08.00+0 DOI: 10.1128/CDLI.11.2.255-260.2004
Copyright © 2004, American Society for Microbiology. All Rights Reserved.
NCI-Frederick/SAIC-Frederick, Frederick,1 Division of Cancer Epidemiology and Genetics, National Institutes of Health, Bethesda, Maryland2
Received 3 November 2003/ Returned for modification 15 December 2003/ Accepted 29 December 2003
|
|
|---|
|
|
|---|
The CRA has proven extremely useful in determining functional CTL responses to cancer, viral infections, and vaccines (12, 13). Despite the disadvantages of having to use radioactive compounds, the requirement for high number of available cells, and time required in assay preparation, this assay still remains in widespread use.
Cytolysis of antigen-specific 51Cr-labeled targets in the standard CRA is determined by the quantification of 51Cr released into the supernatant collected from CTL cultures by using traditional gamma counters. Historically, 51Cr has been considered a gamma emitter, and until the advent of multidetector microplate scintillation counters, the use of gamma counters was favored as the counting method. Considering that 51Cr actually emits more beta particles than gamma particles, the utilization of the multidetector beta counter to quantitate 51Cr release presents an attractive alternative over the traditional gamma counter (4, 10). Simplified procedures utilizing benchtop microplate beta scintillation counters offer several advantages over the traditional method with regard to time and sample management (4, 10). In these procedures, culture supernatants are directly transferred to microplates retaining the original 96-well microplate format. Multiple wells are counted simultaneously, reducing the overall counting time (4, 10).
Several studies use one of these three systems, but there is no published literature that systematically compares the results obtained from each of the three instruments. This comparison is important for laboratories that may have access only to a benchtop microplate beta counter and want to initiate the use of CRA to measure CTL activity against cancer-specific or virus-specific antigens.
Here, we investigated the utilization of two different benchtop microplate beta counter instruments, the Microbeta Trilux microplate scintillation beta counter and the Packard TopCount counter, to quantitate 51Cr release from cytotoxicity assays run in parallel using influenza virus-infected Epstein-Barr virus (EBV)-transformed B-lymphoblastoid cell line (B-LCL) targets, the Daudi and K562 cell lines. 51Cr release was measured with the Microbeta Trilux scintillation counter by using a 96-well microplate format with either a liquid scintillant (flexible beta plates) or a solid scintillant (Lumaplate). Only the Lumaplate format was used to measure cytolysis with the TopCount instrument. These results were then correlated with data collected by traditional gamma counting from an assay run in parallel.
|
|
|---|
Generation of EBV target cells. Cryopreserved PBMC were thawed, washed, and resuspended in CM plus 20% FBS (CM-20) at 6 x 106/ml for the generation of EBV-transformed B-LCL. To each well containing 3 x 106 cells, 0.5 ml of EBV supernatant obtained from B95.8 marmoset cell line (American Type Culture Collection, Manassas, Va.) and 2 µg of anti-CD3 monoclonal antibody (Pharmingen, San Diego, CA) per ml were added (9) EBV-transformed B-LCL were cryopreserved until needed. B-LCL were thawed and placed in CM plus 10% FBS (CM-10) for 3 days prior to CRA.
Preparation of targets for CRA. Cultured EBV-transformed B-LCL were washed in CM and counted, and viability (>90%) was determined. B-LCL (2 x 106 cells) were incubated in a 200-µl volume with either CM (control target) alone or a 1:4 final dilution of influenza virus (American Type Culture Collection) at 37°C in 6% CO2 for 1 h, with gentle agitation every 20 min (1). B-LCL were washed and labeled with Na251CrO4 (100 µCi/106 cells; Perkin-Elmer, Boston, Mass.) for 2 h at 37°C in 6% CO2 with gentle agitation every 30 min. 51Cr-labeled B-LCL were washed three times in CM-10 and adjusted to 5 x 104 cells/ml. Cell viability was determined by trypan blue exclusion.
In some experiments (n = 4), K562 cells were used to determine NK activity and Daudi cells were used as targets for LAK activity (4, 7). PBMC cultured in the presence of interleukin-2 (IL-2; 5 IU/ml; Hoffman-LaRoche, Basel, Switzerland) were used as effectors for the NK cell and LAK assays. An aliquot of 2 x 106 cells was removed from each cell line, washed, and labeled with Na251CrO4 (100 µCi/106 cells; Perkin-Elmer) for 2 h at 37°C in 6% CO2 with gentle agitation every 30 min. 51Cr-labeled Daudi and K562 cells were then washed three times in CM-10 and adjusted to 5 x 104 cells/ml. Cell viability (>90%) was determined by trypan blue exclusion.
Generation and preparation of antigen-specific CTL effector cells. Cryopreserved PBMC were thawed, washed, resuspended in CM-20 at 3 x 106 cells/ml, and incubated with influenza virus at a final concentration of 1:100, as previously described (1). PBMC incubated in CM alone were used as control effectors. Cultures were incubated at 37°C in 6% CO2 for 7 days. On day 3, IL-2 was added to each well at 5 IU/ml. Cultured PBMC were harvested on day 7 and resuspended in CM-10 at 5 x 106 cells/ml. Viability was ascertained as described above.
CRA. Autologous EBV-transformed B-LCL were used as target cells for the influenza virus-specific CTL assays. Equal volumes of target (5 x 104 cells/ml) and effector cells were added to triplicate wells of 96-well tissue culture plates, and 1:2 serial dilutions of effectors were made, producing effector-to-target (E:T) ratios of 100:1, 50:1, 25:1, and 12.5:1 (9). Spontaneous release of 51Cr was assessed by the incubation of target cells in medium alone, and maximum release of 51Cr was determined by the incubation of target cells in 0.1% Triton X-100 (Sigma-Aldrich). CRA plates were set up in duplicate to run traditional gamma counting in parallel with the other counting platforms. After a 4-h incubation of the effector cells with the target cells at 37°C in 6% CO2, supernatants were collected following brief centrifugation.
For the traditional measurement of 51Cr release, the supernatant was collected from one of the duplicate CRA culture plates into filters of supernatant collection system harvesting frames (Molecular Devices Skatron, Sunnyvale, Calif.) and transferred to polystyrene tubes to be counted with the LKB 1272 Clinigamma counter (Wallac). This method (LKB) was run in parallel with other methods adapted from manufacturers' recommendations.
The supernatant from the second duplicate CRA culture plate was collected and transferred to a new 96-well microplate (supernatant plate), and aliquots were taken from this plate for each of the following procedures. For the methods using the Lumaplates (Perkin-Elmer), an aliquot of 50 µl was directly transferred from the supernatant plate to the Lumaplate and allowed to air dry in a hood overnight. Lumaplates are opaque-walled 96-well microplates containing a thin layer of solid scintillant on the bottom of the wells. Plates were sealed and counted in the Packard TopCount counter (Perkin-Elmer) for 1 min per well with dual detectors using a 0-to-256 channel window with the instrument protocol set for solid scintillant conditions (TopCount method). The isothermal counting chamber of the instrument was set at 19°C. The same Lumaplates were then counted on the Microbeta Trilux instrument (Perkin-Elmer) (Lumalux method) for 1 min per well with dual detectors using a 5-to-170 channel window. All detectors were normalized according to manufacturers' suggestions.
The Lumaplate and tube transfer methods were then run in parallel with the method using the 96-well flexible beta plates (Perkin-Elmer) by using a nonvolatile liquid scintillant, Optiphase SuperMix (Perkin-Elmer). Flexible beta plates are constructed of polyethylene terephthalate and contain a printed gridline pattern on the surface to separate each well. An aliquot of 25 µl of CRA supernatant was transferred from the supernatant plate to the flexible beta plate containing 150 µl of Optiphase SuperMix. The plate was sealed, shaken vigorously for 15 min, and then counted on the Microbeta Trilux instrument as described above (Trilux method).
Optical cross talk was minimized by the use of the opaque-walled Lumaplates and the printed gridlines on the flexible beta plates. Isotopic cross talk was corrected for by the software protocol for each instrument. However, for chromium, the beta events are concentrated in one part of the spectrum and the high-energy gamma events that give rise to the cross talk problem are located at the higher end of the spectrum. The channel window setting of 5 to 170 minimizes the effect of cross talk. The cross talk of 1 µCi of 51Cr in flexible beta plates run in the Microbeta Trilux instrument was 0.06%.
Percent specific lysis for 51Cr release was determined with the following equation: [(experimental 51Cr release - spontaneous 51Cr release)/(maximum 51Cr release - spontaneous 51Cr release)] x 100.
All experiments were performed in triplicate wells. Results are expressed as mean percent specific lysis ± standard deviation or mean counts per minute of triplicate wells. Percent spontaneous release in all of the assays was
12.4% for any target. Background levels (mean counts per minute ± standard deviation) for empty wells and tubes (n = 22) were 62.8 ± 16.2 for the LKB method, 3.7 ± 4.0 for Trilux, 2.0 ± 1.2 for TopCount, and 2.2 ± 9.1 for Lumalux.
Statistical analysis. Data obtained from other counting platforms were correlated with data collected by the traditional gamma counting procedure using the Pearson correlation coefficient. Spearman correlation coefficients yielded nearly identical findings and are therefore not reported. The nonparametric Mann-Whitney test was used to determine statistical differences between counting methods. A P value of <0.05 was considered significant.
|
|
|---|
![]() View larger version (25K): [in a new window] |
FIG. 1. Comparison of CTL responses of influenza virus-stimulated effectors to uninfected (A) or influenza-infected EBV transformed B-LCL (B) targets evaluated by the quantification of chromium-51 release using the benchtop microplate beta counters Microbeta Trilux and TopCount against traditional gamma counting. Microbeta Trilux readings were taken from both solid (Lumalux) and liquid (Trilux) scintillation formats. Data are CTL responses from seven individual experiments, as specified in Materials and Methods.
|
|
View this table: [in a new window] |
TABLE 1. Spontaneous and maximum release ranges for uninfected and influenza virus-infected targets as well as NK and LAK targets for each counting method
|
The net percent specific lysis of influenza virus-specific responses relative to the medium control ranged from 22.3% at an E:T ratio of 100:1 to 18.7% at an E:T ratio of 12.5:1 with Trilux, 16.3 to 12.2% with TopCount, and 18.1 to 14.5% with Lumalux, compared to 20.9 to 18.0% with LKB.
LAK and NK activity (n = 4) was measured against Daudi (Fig. 2A) and K562 (Fig. 2B) cell lines by the four methods. LAK activity against Daudi targets measured with the LKB method had a mean range of 59.5% lysis at an E:T ratio of 100:1 to 31.2% lysis at an E:T ratio of 12.5:1 and 69.9 to 42.5% for NK activity against K562 targets. The TopCount method showed mean lytic ranges of 50.0 to 26.8% and 51.6 to 30.5% for LAK and NK activity, respectively. Trilux produced ranges of 63.1 to 36.0% for Daudi and 70.7 to 43.6% for K562 cell lysis. Mean ranges of 52.3 to 29.7% lysis for Daudi and 54.9 to 35.0% lysis with K562 targets were observed with the Lumalux method. Both NK and LAK activity corresponded well with the trend of lytic activity observed with the influenza virus-infected targets. Results demonstrated higher percentages with the LKB and Trilux methods than with the TopCount and Lumalux methods, but all followed similar patterns of titration. However, statistically significant differences were observed only between LKB and TopCount methods against K562 targets at E:T ratios of 100:1 to 25:1 (P = 0.03) and between LKB and Lumalux at ratios of 100:1 and 50:1 (P = 0.03). No significant differences were seen with the Daudi targets.
![]() View larger version (26K): [in a new window] |
FIG. 2. Lytic activity of cultured PBMC in the presence of IL-2 assayed for LAK activity against Daudi targets (A) and for NK activity against K562 targets (B) measured from cytotoxicity assays. Lytic activity from 51Cr released into cell culture supernatants was measured with the microplate readers Microbeta Trilux and TopCount and by traditional gamma counting (LKB). Microbeta Trilux readings were done in the presence of solid (Lumalux) and liquid (Trilux) scintillation cocktails. CTL responses are means from seven experiments.
|
0.957 at 100:1,
0.950 at 50:1,
0.984 at 25:1, and
0.956 at 12.5:1.
![]() View larger version (16K): [in a new window] |
FIG. 3. Correlation of specific lysis of influenza virus-stimulated effectors against influenza virus-infected and control uninfected targets obtained by traditional gamma counting with values using flexible plates read with the Microbeta Trilux counter (A) and Lumaplates read with the TopCount (B) or Microbeta Trilux (C) counter.
|
![]() View larger version (16K): [in a new window] |
FIG. 4. Correlation of mean counts per minute of influenza virus-stimulated effectors against influenza virus-infected and control uninfected targets obtained by traditional gamma counting with those obtained from the Microbeta Trilux instrument using a solid (Lumalux) and a liquid (Trilux) scintillant and from the TopCount instrument using solid-scintillant Lumaplates.
|
0.975; counts per minute, r
0.962) between instruments for the methods used, as did LAK activity against Daudi targets (specific lysis, r
0.979; counts per minute, r
0.990). |
|
|---|
Here, we compared influenza virus-specific CTL, LAK, and NK cell responses assessed by chromium-51 release measured with three different instruments using either transfer tubes or 96-well microplates containing either a solid or liquid scintillant. Cytotoxicity results obtained with the different benchtop microplate beta counters correlated very well between instruments and with the traditional gamma counting system. However, the traditional LKB method produced the highest percent specific lysis with the uninfected and influenza virus-infected EBV targets as well as the K562 targets. The Trilux method showed results very similar to those of the LKB method and higher lytic activity with the Daudi targets. With the LKB method, the CRA traditionally yields a 200-µl volume of supernatant that is collected into the harvesting filters. This volume is 4 to 8 times higher than the amount required for each of the alternative methods which demonstrated similar patterns of lytic activity. This advantage over the LKB method allows a reduction in sampling size requirements. In addition, the use of the microplate format to count chromium release in cytotoxicity assays eliminates potential errors associated with tube manipulation. Samples prepared in the 96-well plate format can be counted directly in that format, ensuring correct positioning of samples, and each plate replaces the need for 96 tubes. Maintaining the microplate format facilitates the possibility of automated liquid sample handling in the cases when liquid scintillation cocktail is used. The benchtop microplate beta counters also allow the utilization of the same equipment for detection of gamma- and beta-emitting isotopes. Some of these counters also have the ability to increase the number of detectors to allow up to 12 wells to be read simultaneously, greatly reducing the overall counting time and optimizing data output.
With increasing concerns over the safety and disposal of liquid scintillation cocktails, solid scintillation counting using Lumaplates represents a further attractive alternative to liquid counting in a microplate format by reducing the amount of mixed radioactive and chemical liquid waste generated. The lower volume size required with benchtop microplate beta counter formats also allows assay cultures to be scaled down, reducing total cell numbers and sampling sizes and minimizing the amount of radioactive waste generated. These features make the TopCount and Lumalux methods suitable for high-throughput settings.
However, the applicability of the microplate methods to the analysis of cytolytic responses to weak inducers of CTL activity or from patients with compromised immune systems needs to be further examined.
In summary, our results indicate that supernatants from CRA can be counted in solid or liquid scintillant with Microbeta Trilux or TopCount Instruments and that each of these methods correlates significantly with the traditional gamma counting method.
We thank Mohammad Ishaq for providing access to the Packard TopCount instrument and Alan Doty and Richard Jones for their technical support and recommendations for the methods outlined here.
|
|
|---|
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||