Clinical and Diagnostic Laboratory Immunology, July 1998, p. 513-518, Vol. 5, No. 4
CERVI-VIROLOGIE, Centre Hospitalier
Universitaire Pitié-Salpêtrière, 75013 Paris,
France
Received 11 August 1997/Returned for modification 1 October
1997/Accepted 19 March 1998
The sensitivity of the enzyme-linked amplified sorbent test (ELAST)
was compared with those of other classic enzyme-linked immunosorbent assays (ELISAs), with or without previous acidic immunocomplex dissociation (ICD), in a series of samples at different stages of human immunodeficiency virus type 1 (HIV-1)
infection. The limit of viral detection of ELAST was assessed with
fresh HIV-1 preparations quantified by reverse transcription-PCR and with the P24 antigen (Ag) Sanofi Pasteur Calibrator containing lyophilized virus. The P24 Ag detection capacity of ELAST was compared
with that of NASBA in samples obtained from infected subjects with less
than 250 CD4+ cells. The results of the present study show
that ELAST was the most sensitive method for detecting P24
Ag compared to classic ELISA and ICD plus ELISA. ELAST was able to
detect 0.5 pg of P24 Ag per ml in a whole virus preparation and the
equivalent of 330 to 1,000 RNA copies/ml of HIV. The rate of detection
of P24 Ag was always higher in subjects with low levels of anti-P24
antibodies. The number of positive results was dramatically enhanced
(from 37% to 94% for subjects with <250 CD4+ cells) when
the incubation period was prolonged from 1 to 16 h. In a third
series of 84 samples (<250 CD4+ cells) tested in parallel,
NASBA yielded 83% of the positive results and ELAST yielded 79%.
Considering the high sensitivity, low cost, simplicity of equipment
(only a plate reader), and possibility for full automation, ELAST
appears to be a promising new tool for measuring viral load, especially
in areas with few resources, in which the procedures based on
molecular biology techniques may be difficult to install.
Several surrogate markers have been
used to monitor disease progression in human immunodeficiency virus
(HIV)-infected individuals (23, 42). These markers
include CD4 count, serum P24 antigen (Ag) levels with or without immune
complex dissociation (ICD) before ELISA, quantitative culture of
peripheral blood mononuclear cells (PBMCs) or plasma, DNA levels in
infected circulating cells, and RNA levels in plasma. A number of
studies have shown that in HIV-infected persons, an increased viral
burden is associated with the degree of immune deficiency and enhanced
risk of perinatal transmission of the virus to the child (19,
38). Moreover, the correlation between the disease stage,
CD4+ cell number, and HIV-1 provirus and virus levels has
frequently been reported (19, 23, 25, 38, 42, 47). Methods
that are used for quantification of circulating HIV-1 include in vitro culture (cell or plasma) (15, 40), PCR of serial dilutions of PBMCs (1), reverse transcription-PCR (RT-PCR) of serial dilutions of plasma or serum (21, 51), and amplification
techniques with no need for serial dilutions (49). Overall,
the results of these approaches indicate that there is a significant
association between levels of viral activity and disease progression.
The measure of circulating HIV load allows for the monitoring of
treatments (3, 31) and permits a better understanding of the
natural history of the infection (11, 34).
HIV p24gag Ag (P24 Ag) has been widely used as a
marker of virus replication. However, only a small fraction of
HIV-infected individuals have measurable circulating P24 Ag levels when
measured by classical immunocapture techniques (enzyme-linked
immunosorbent assay [ELISA]) (14, 33).
The P24 Ag level has been shown in different studies to correlate with
disease progression (8, 16, 23, 32, 36), and it was further
suggested that P24 Ag may be used as a criterion for entry into
clinical trials and subsequently for monitoring antiviral therapy
(24, 29, 43). Treatment of serum samples with acid (26,
28) or base (41) can dissociate P24 Ag complexed with
antibodies and increase the rate of positivity (ICD) (31, 35,
39). However, it is no longer accepted that P24 Ag ELISAs, even
after ICD, meet the goals of sensitivity required for reliable surrogate markers (12). Thus, the quantification of the HIV viral load based on molecular biology techniques (MBTs) (RT-PCR, NASBA
[Organon Teknika], and branched DNA) has become the yardstick for the
follow-up of HIV infection. Nevertheless, one of the major concerns
which limits the widespread use of MBTs is intra- and interlaboratory
variability (technical problems, nondetectable genotypes, etc.).
Moreover, the cost of setting up the required facilities to avoid
contamination, the price of materials and reactants, and the time
required for comprehensive training of staff in MBTs, etc., curtail
even further the utilization of MBTs, especially in areas with reduced
resources. Therefore, there is a crying need to evaluate and validate
affordable techniques to assess HIV viral load.
Amplification of signals of bound proteins with a catalyzed-reporter
deposition technique (enzyme-linked amplified sorbent test [ELAST])
(DuPont de Nemours-NEN, Paris, France) enhances levels of detection of
P24 Ag (9). Schüpbach et al. (45) recently
showed that disruption of immunocomplexes by heating 1/6-diluted plasma
prior to ELAST detected HIV expression as sensitively as RT-PCR.
The aim of this work is to compare the results obtained with ELAST with
those of other ELISAs used currently to detect P24 Ag in a series of
human samples at different stages of HIV-1 infection. Sensitivity was
gauged with (i) a fresh HIV-1 preparation quantified after serial
dilution by RT-PCR and NASBA and (ii) the P24 Ag Sanofi Pasteur
Calibrator (lyophilized virus recognized as standard by
the French health authorities). Finally, in a different series of blind
samples (plasma obtained from HIV-infected people), ELAST was compared
to NASBA.
Samples.
In all of the studies presented here, the tests
were performed with blind samples. The first two series of sera
(classic ELISA versus ELAST) were obtained from different subjects
living in France, in which the dosage of P24 Ag was prescribed before
MBTs were available for routine HIV viral load determinations. The sera
were kept at
1071-412X/98/$04.00+0
Copyright © 1998, American Society for Microbiology. All rights reserved.
Comparison of an Amplified Enzyme-Linked Immunosorbent Assay with
Procedures Based on Molecular Biology for Assessing Human
Immunodeficiency Virus Type 1 Viral Load
![]()
ABSTRACT
Top
Abstract
Introduction
Materials & Methods
Results
Discussion
References
![]()
INTRODUCTION
Top
Abstract
Introduction
Materials & Methods
Results
Discussion
References
![]()
MATERIALS AND METHODS
Top
Abstract
Introduction
Materials & Methods
Results
Discussion
References
20°C for more than 12 months. The clinical status and
immunological status of these subjects were monitored in public
hospitals. Since genotyping or subtyping is not performed during
routine follow-up, no data on the HIV-1 subtypes were available (in
France, clade B is largely in the majority).
80°C. NASBA and ELAST were performed under
routine conditions.
P24 standard.
The standard (Sanofi-Diagnostics Pasteur)
approved by the Agence du Medicament (France)
containing
lyophilized HIV-1 (BRU strain) grown in CEM cells,
purified by ultracentrifugation, inactivated and diluted in human
serum, and negative for hepatitis B surface Ag, hepatitis C viral
antibodies, and HIV-1 and HIV-2 antibodies
was used for comparative
tests of sensitivity. For each experiment, calibration curves were
plotted with dilutions of the standard in a pool of HIV-1- and
HIV-2-negative sera, ranging between 0.1 (100 fg) and 64 pg/ml.
Testing for P24 Ag. Samples were tested by the immunocapture ELISA (HIV-p24 Core Profile ELISA; Du Pont de Nemours) according to the manufacturer's directions. Ag-antibody complexes (immunocomplexes) were disrupted (ICD) by using low-pH solutions (1 volume of sample plus 1 volume of glycine reagent with 1 volume of Tris for neutralization).
ELAST. Circulating P24 Ag was detected with the immunocapture ELISA (HIV-p24 Core Profile ELISA; Du Pont de Nemours) followed by amplification of the bound antigen (5, 7). For running the tests, 20 µl of 5% Triton, 100 µl of samples or controls, and 100 µl of a glycine solution (pH 2) were incubated for 90 min at 37°C for ICD. Neutralization was performed with 100 µl of Tris buffer for 10 min at room temperature. (Only in a small number of tests [<5%] did we experience gelling after ICD. When that happened, samples were strongly agitated with the micropipettes.) It was always possible to perform the tests on the volume outside the gel (>150 µl in all cases).
One hundred fifty microliters was then transferred to the ELISA plate (HIV-1 p24 Core Profile Add Kit; Du Pont) and incubated for 1 h or overnight (16 h) at room temperature. (After each of the incubations during ELAST, a specific washing program was applied. Plates were washed with 300 µl of DuPont's washing solution nine times.) After washing, 100 µl of detector antibody was added, and plates were incubated for 1 h at 37°C. After washing (nine times), horseradish peroxidase (HRP)-S was added (150 µl of HRP-S concentrate in 10 ml of HRP dilutant), and this mixture was incubated for 30 min at room temperature. After washing, 100 µl of biotinyl-tyramide (BT) was added (100 µl of BT concentrate in 10 ml of BT-diluent at a 1/2 ratio). HRP converts phenol derivatives to free radicals and catalyzes the activation of the phenolic groups of BT, resulting in covalent binding to protein-blocking molecules on the solid phase. Plates were incubated for 30 min at room temperature and washed, and 100 µl of HRP-ELAST solution was added (20 µl of HRP-ELAST in 10 ml of ELAST dilutant for HRP). After incubation (30 min at room temperature) and washing, 100 µl of OPD was added, and the reaction was stopped after 15 min (100 µl of HCl [1 N]). Optical densities (ODs) were determined at 490 or 620 nm, and concentrations were calculated from the calibration curves plotted with the standard at concentrations between 0.10 and 64 pg/ml. For each experiment, cutoff values were calculated as the mean OD for the negative controls + 2 standard deviations. Six negative controls were randomly included in each series of 96 determinations.Quantification of HIV-1 in a supernatant of infected cells.
RT-PCR was performed to titrate the internal standard, which was
prepared as follows. Supernatants of HIV-1-infected human lymphocytes
were clarified for 10 min at 1,000 × g and
ultracentrifugated for 10 min at 350,000 × g in a
Beckman TL-100 centrifuge. The pellet was diluted in 100 µl of
guanidium thiocyanate buffer, and 1/10 acetic acid, 1/2 volume of
phenol, and 1/2 volume of CHISAM (49 volumes of Cl3CH-1
volume of isoamyl alcohol) were then added. The aqueous phase was
transferred to a new tube, and 1 µl of glycogen and 1.5 volume of
isopropanol were added. RNA was allowed to precipitate overnight at
20°C and centrifuged for 20 min at 11,000 × g. The
pellet was then treated with DNase (10 U for 10 min at 20°C). The
enzyme was inactivated at 100°C for 2 min. RT of RNA (18,
30) was performed with a volume of 20 µl containing 50 mM
Tris-HCl (pH 8.3), 75 mM KCl, Cl2Mg, 0.5 mM (each)
deoxynucleoside triphosphates, 10 mM dithiothreitol, and 200 U of
Moloney murine leukemia virus reverse transcriptase (MoMuLVRT) (Gibco
BRL), 20 U of RNase inhibitor (Boehringer), and 10 pmol of Pol 4 antisense primer. MoMuLVRT (Gibco) was added after RNA denaturation at
65°C for 5 min. PCR was performed with Taq polymerase
(Beckman) and 2 µl of cDNA. Amplification was performed with 38 cycles (92°C, 55°C, 72°C) with two sets of primers bracketing the
SK 38 and 39 and the Pol 3 and 4 fragments (18, 30). PCR products were electrophoresed in agarose gels, transferred to nylon
membranes, and hybridized with radiolabelled internal probes (Pol 3 and
4 and SK 19). RNA content was assessed according to endpoint dilution
methods (2, 18, 30). (The titer of the standard preparation
was 1,000 RNA copies/ml.) When NASBA and RT-PCR kits became available
for routine use, we retested the standard preparation with them
according to the manufacturers' indications (at respective titers of
700 and 900 RNA copies/ml).
NASBA. Tests were performed with 500 µl of plasma. Nucleic acids were released into a chaotropic agent and bound to silicon dioxide particles under high-salt conditions (10). Solid particles were washed, and nucleic acids were eluted and amplified by interaction of the enzymes avian myeloblastosis virus reverse transcriptase, RNase H, and phage T7 RNA polymerase, together with two primer sets selected from the HIV-1 gag region (49). Three synthetic RNAs of known high, medium, and low concentrations (undiluted internal calibrators), respectively, were added to the lysis buffer containing the released nucleic acids. Detection of HIV RNA in samples was performed by electrochemiluminescence (27). The results of NASBA were expressed in RNA copies per milliliter.
Determination of anti-P24 antibodies. Tests were performed according to the Abbot HIV P24 recombinant DNA kit's directions. The Abbott anti-P24 software was integrated with the ELISA plate reader. Sera with ODs less than the cutoff at a dilution of 1/5 were considered negative.
| |
RESULTS |
|---|
|
|
|---|
Table 1 shows the results obtained
with the first series of 279 samples. Classic ELISA was able to detect
only 8 out of 114 samples positive for the P24 Ag when the sera were
obtained from persons with less than 250 CD4+ cells. When
acidic ICD was performed before ELISA, 26 samples became positive. The
amplification of the signals with no prior ICD (classic ELISA plus
amplification) showed 38 positive results, and ICD plus amplification
(ELAST) showed 42 positive results. Even after dilution (for ICD, sera
were diluted 1/3 with the acidic and the neutralizing solutions), ELAST
showed the highest rates of positivity for all of the samples
independently of their CD4+ cell count ranking. In samples
obtained from subjects with more than 500 CD4+ cells
in
which classic ELISA did not detect any positivity
a dramatic increase
in sensitivity was obtained with ELAST, with 11 positive samples out of
30 (37%). For sera obtained from subjects with CD4+ cell
counts between 250 and 500, ELAST was also the most sensitive method,
with positive results for 33% of the samples. Global analysis of the
results of the first series of 279 HIV+ sera (in which at
least 40% were obtained from subjects in an advanced stage of
infection), indicates that only 13 (4.2%) samples were positive
by classic ELISA. After ICD, 22% of the samples became positive. When
amplification was performed after ELISA, but with no previous
ICD, 24% of all of the samples were positive. ICD previous to ELISA
and BT amplification (ELAST) increased the rates of detection to 36%.
|
Table 2 shows the results of the first series of 279 samples stratified into two groups: positive and negative anti-P24 antibodies. The levels of Ag detection for all methods (classic ELISA, ICD plus ELISA, classic ELISA plus amplification, and ELAST) were significantly higher in samples with no anti-P24 antibodies. ELAST showed the highest level of antigen detection (19% of the positive samples and 85.7% of the negative samples). ICD with no amplification was less sensitive (10% and 55.7%, respectively). Classic ELISA did not detect the P24 Ag in any of the samples obtained from subjects with detectable levels of anti-P24 antibodies.
|
Acidic ICD enhanced the detection levels for P24 Ag from 0 to 10% in the group of anti-P24+ subjects. With no prior ICD, BT amplification enhanced the level of detection up to 11.5%. The coupling of ICD with amplification (ELAST) in the anti-P24+ group enhanced detection levels to 19.6%.
In a second series of 100 samples (CD4+ count of <250),
ELAST was performed after 1 or 16 h of incubation. P24 Ag was
positive for 37% of the samples after 1 h of incubation, but the
rate dramatically increased to 94% when the incubation period
was prolonged up to 16 h. Out of 100 samples tested, 62 samples
were negative and 1 was borderline (sample ODs
10%, which was the
cutoff value) when the Ag immunocapture on the microplates was
performed according to the manufacturer's indications (1 h). When the
incubation was prolonged up to 16 h, only five samples were
negative and one sample was borderline. Forty samples from
HIV-seronegative subjects were also run in triplicate, and they were
all negative for ELAST performed after 1 or 16 h of incubation.
For ELAST (16 h), reproducibility was 100% for negative samples and 100% for all of the samples containing the equivalent of >1 pg/ml. For 15 samples (0.25 to 1 pg/ml) tested repeatedly (four times), reproducibility was 95% and 57 determinations showed the same values (picograms per milliliter). Intralaboratory variability for retesting of the samples with reactants from different lots (picograms per milliliter) was always <20%. With different lots of reactants retested at different dates, linearity ranged between 0.5 and 16 pg/ml.
In a preliminary study, we compared detection limits for classic ELISA, ELAST with 1 h of incubation, and ELAST with 16 h of incubation. According to the curve performed with the standard, they were, respectively, 20, 2, and 0.25 pg/ml. In the present study, Table 3 shows the results obtained with a supernatant of a fresh HIV-1 preparation serially diluted. This preparation was titrated by RT-PCR and NASBA, and the results with the aliquot used as a standard were, respectively, 1,000 and 800 RNA copies/ml. Classic ELISA was able to detect the equivalent of 20,000 viral RNA copies. ELAST was able to detect the equivalent of 2,500 viral RNA copies with an incubation time of 1 h. When the incubation was prolonged for 16 h, ELAST detected the equivalent of 1,000 viral RNA copies. These results were reproducible in four different tests with determinations performed each time in triplicate. Since before antigen capture, all of the samples were diluted 1/3 for acidic ICD, the signal detection limit of ELAST (16-h incubation) may be adjusted down to the equivalent of at least 1,000 viral RNA copies/ml.
|
Table 4 shows the results obtained with a third series of 84 samples (plasma) obtained from persons with less than 250 CD4+ cells (range, 50 to 250), tested in parallel with NASBA and ELAST. From these 84 samples, 10 were negative by the two techniques. NASBA yielded 83% positive results, and ELAST yielded 79% positive results. Four samples were found positive by ELAST and negative by NASBA. (Three of these four samples showed weak ODs, corresponding to concentrations below 1 pg/ml according to the standard curve.) Eight samples were positive by NASBA but negative by ELAST, and one sample (ELAST negative, repeated three times) had more than 40,000 copies of HIV RNA according to NASBA. Two samples containing between 800 and 2,000 RNA copies and five containing between 2,000 and 40,000 RNA copies were not detected by ELAST. Six samples out of eight with viral load values (NASBA) of <2 × 103 RNA copies/ml were detected by ELAST (results ranged between 0.1 and 4 pg/ml). ELAST rates of positivity were 85% in the NASBA category >2 × 103 and <40 × 103 RNA copies/ml (44 samples) and 95% when viral load values ranged between >40 × 103 and <200 × 103 RNA copies/ml (20 samples). All of the samples with values of >2 × 105 RNA copies/ml were detected by ELAST.
|
The plasma samples tested in the third series of comparative
studies were frozen at
20°C and thawed for retesting
(triplicate) 1 month later. Intra- and intertest variabilities
for ELAST (16 h) were <20% (picograms per milliliter).
| |
DISCUSSION |
|---|
|
|
|---|
No laboratory test (whether virological or immunological) is able today to fulfill the expected requirements of correlation with progression, change if the patient improves or worsens, variation with successful treatments, direct association with HIV natural history, intra- and interlaboratory reproducibility, facilities for setup and routine running (i.e., not requiring sophisticated structures or larger surfaces), comparable cost to other laboratory tests, and affordability to the largest number of infected people around the world.
HIV-1 quantification by the MBTs RT-PCR, NASBA, and branched DNA (20, 37, 46) is gaining more and more ground in determining therapeutic performance without having to consider survival as the endpoint. However, because they are expensive and technically complex, MBTs are routinely used only in a few specialized centers (13, 50). In light of the above criteria, the search for reliable, inexpensive, easy-to-perform tests to determine HIV viral load has been a major concern for virologists.
The results of the present study indicate that ELAST (ICD plus ELISA coupled with the BT amplification system) is the most sensitive method for detecting P24 Ag, compared to classic ELISA, ICD plus ELISA, and amplification with no previous ICD.
Schüpbach et al. (45) reported in a previous study that the rate of positivity was enhanced from 27 to 53% when sera were treated with acids for ICD and to 68% when the immunocomplexes were disrupted by heating the samples (1/6 diluted) for 5 min at 100°C.
In the present study, acid ICD enhanced the detection levels for P24 Ag from 0 to 10% in the group of anti-P24+ subjects. Even with no prior ICD, BT amplification also enhanced the level of detection up to 11.5%. The coupling of ICD with amplification (ELAST) in the anti-P24+ group enhanced detection levels to 19.6%.
Schüpbach et al. (45) indicated that overall rates of
detection for samples (diluted 1/6) obtained from HIV-infected people were 68% for sera and 78% for plasma. In the present study, the detection rate is almost identical to that obtained by Schüpbach et al. (45) for sera not treated for Ag-antibody
dissociation. Nevertheless, the rates of positivity obtained by
Schüpbach et al. were higher, both when immunocomplexes were
dissociated by acids and when they were dissociated by heat. Positive
results for sera in our study versus that of Schüpbach's with
sera and versus that of Schüpbach's with plasma for
CD4+ cell counts of 200 to 500 are 33, 49.4, and 63%,
respectively. The rates for the group with a CD4+ cell
count < 250 are 37, 61, and 74%, respectively. Since for the
group of samples with CD4+ counts > 500, the results
were, respectively, 37 and 5.6%, we cannot explain the differences
observed in the first two groups just by the long storage of our series
of samples at
20°C.
In the present study, the levels of P24 Ag detection for all methods
(ELISA, ICD plus ELISA, classic ELISA plus amplification, and ELAST)
were significantly higher in samples with no anti-P24 antibodies. When
comparing the results obtained with the samples for each technique, for
anti-P24 positive versus anti-P24 negative samples, ELAST showed the
highest level of antigen detection (16.2% of the positive samples and
85.7% of the negative samples). ICD with no amplification was less
sensitive (9.6% and 55.7%, respectively). Moreover, the lack of
detection of P24 Ag is not only due to the entrapment of the viral
particles or proteins by the circulating antibodies, but may reflect
low levels of viral replication and/or release of virions from
reservoirs. Viral replication assessed by the positivity of a freely
circulating viral Ag may therefore be one of the consequences of a
reduction of immune functions, since for all of the techniques tested
here, the highest rates of P24 antigen detection were obtained in
subjects with low (or undetectable) levels of anti-P24 antibodies
(anti-P24
). Recently, Binley et al. reported results
along the same lines and concluded that the disappearance of anti-Gag
antibodies during infection is highly unlikely to be due to immune
complex formation (6).
The first series of sera (Table 1), from persons with less than
250 CD4+ cells, showed a rate of positivity for ELAST never
higher than 37% (anti-P24+ and anti-P24
together). This value is lower than that expected from previous reports in which the positivity rate for sera obtained from subjects with CD4+ cell counts < 200/ml was 60% (45).
To enhance the level of detection, we performed kinetic studies with
samples with ODs higher than the average of the negative control ODs
but lower than the threshold (average negative control ODs + standard deviations). We carried out these studies with different
incubation times (4, 8, 16, 24, 36, and 48 h). More than 50% of
the samples became positive when the incubation period reached 16 h (plateau at 16 h). When standard (whole virus preparation)
dilutions (four tests in triplicate in each experiment) were randomly
tested, the P24 Ag detection limit for classic ELISA was never under 16 pg/ml. ELAST made it possible to detect 2 pg of P24 Ag per ml after
1 h of incubation and 0.5 pg/ml after 16 h (with 100% of the
samples containing 0.5 pg of the whole viral preparation per ml
positive in all of the experiments).
Böni et al. (9) reported that detection rates for samples obtained from treated patients during a median observation time of 25 weeks (Indinavir at 800 mg three times daily in addition to prior antiretroviral treatment) amounted to 75.6% for amplified ELISA and to 73.6% for RT-PCR (detection limit of 0.2 pg/ml). Böni et al. indicate that individual antigen levels correlated well with RT-PCR (r = 0.714). By analyzing their results in terms of group means at different time points, the average treatment induced changes in viral RNA and was closely matched by changes in viral antigen. Van Damme et al. (48) compared the abilities to measure viral load in 107 samples from 25 subjects by RT-PCR (Monitor; Roche) and NASBA and obtained a correlation coefficient on the same order as that obtained when RT-PCR and amplified P24 Ag are compared (r = 0.688). Coste et al. (17) showed in a comparative study carried out with 60 plasma samples from HIV-infected people, a concordance of positive results of 80.4% between RT-PCR and NASBA. No correlation between NASBA or RT-PCR and rates of P24 Ag positivity was found. This lack of correlation (r < 0.3 for P24 Ag versus NASBA or versus RT-PCR or versus branched DNA) can be explained by the fact that for P24 Ag determinations, no amplification of the signals was performed and the cutoff value was set as high as 30 pg/ml.
In the present study, NASBA and ELAST testing of plasma from persons with less than 250 CD4+ cells (range, 50 to 250) yielded, respectively, 83% (>400 copies/ml) and 79% positive results. Intriguingly, one sample that was ELAST negative (repeated three times) had more than 20,000 copies of HIV RNA according to NASBA, and one sample that was ELAST positive (4 pg/ml) was NASBA negative. This sample was obtained from an individual included in a double-masked clinical trial (zidovudine plus D4T or placebo) just a few weeks before. The half-life of the released P24 proteins may be longer than RNA release after lysis from intact virions, and we should expect that the reduction in circulating P24 antigen levels may be delayed after initiation of an efficient antiviral treatment. Böni et al. (9) reported the same type of results, indicating that in a reduced number of subjects after starting antiretroviral treatments, the decrease in P24 Ag is slower than that of viral RNA or particle-associated RT activity. In the present study, four samples (4.25%) were repeatedly positive for ELAST and repeatedly negative for NASBA. No inhibitors of NASBA were detected in the plasma (controls were included for extraction amplification and quantification procedures), and technical problems were not reported during the running of these tests in the routine laboratory. Coste et al. (17) reported that NASBA is able to detect most HIV-1 subtypes, but not subtype G. For the samples studied here, no data on HIV-1 subtypes that could explain these discrepancies are available.
Moreover, in a 4-year period, the Amplicor kit was used in a United Kingdom reference laboratory for the detection or confirmation of HIV-1 infection (4). This study showed that the Amplicor assay has a sensitivity of 93.7% for adults and children over 2 years. For specimens from young children (under 2 years of age), sensitivity was 75%. The negative results were assumed to be false negative on the basis of clinical data, serological markers (including P24 Ag), and/or results for previous follow-up specimens.
Differences between MBTs and P24 Ag were also recently reported for different HIV-1 strains: significantly lower RT-PCR (Amplicor) loads were measured for most GagA, GagB and GagF strains, whereas NASBA detected lower copy numbers in the GagH strain, GagH/EnvG recombinant strain, and in one HIV-1 group M strain. HIV-1 group O strains were not detectable by RT-PCR or NASBA but were detectable by the quantification of the P24 Ag (22).
Indeed, the rate (4.2%) of negative results yielded in the present study by NASBA with samples obtained from people in advanced stages of HIV-1 infection (similar rates were reported by others [9]) should be considered as a warning not to use only one method for determining viral load, even with sensitized MBTs (44). We can therefore conclude that if for any subject no previous viral load data are available, the nondetection of circulating viral genomes needs to be confirmed by another test (MBTs and/or ultrasensitive methods able to detect small amounts of circulating viral Ags).
It would appear, from the findings presented above, that amplified ELISA technologies have a role to play as a backup for MBTs. There is, in fact, a definitive advantage in conducting viral load checks at shorter intervals than is the case at present with MBTs. In asymptomatic subjects, for example, frequent checking would permit early detection of viral replication and, consequently, timely treatment. Also, in treated subjects, frequent viral load evaluations may warn of possible therapeutic failure.
The problem here involves both money and facilities: only a few
institutions
not to mention individuals
can afford to provide regular
follow-up tests with MBTs. ELISA technologies, however, are
significantly cheaper than MBTs (less than 10 times the cost for
samples tested with ELAST duplicate tests versus a single test with
RT-PCR, NASBA, or branched DNA). Those laboratories around the
world already using the standard ELISA may easily add signal
amplification techniques, providing them with the added advantage of
automatic processing of large numbers of samples.
If the detection threshold of 1,000 RNA copies/ml (signal produced by the equivalent of 330 copies/ml, considering a 1/3 dilution for ICD) found in the present study is confirmed by others, ELAST (acid-ICD or heat-ICD plus antigen immunocapture plus BT amplification) may become a new tool for routine determination of viral load in cases such as (i) early seroconversions, (ii) the paucisymptomatic phase of the HIV infection, (iii) mother-child HIV-1 transmission, and (iv) whether to start or wait and add or change treatments. However, comparative kinetic studies measuring decay in RNA and in P24 Ag should be performed, because ELAST (16-h incubation) might lead to erroneous conclusions in cases in which antiretroviral therapies induce very rapid reduction in viral replication.
| |
ACKNOWLEDGMENTS |
|---|
We thank Michel Segondy, Laboratoire de Virologie, Hôpital Saint Eloi, Montpellier, France, for help with sample collection and Wendy Kay Johnson for proofreading assistance.
| |
FOOTNOTES |
|---|
* Corresponding author. Mailing address: CERVI-VIROLOGIE, C.H.U. Pitié-Salpêtrière, 87 Boulevard de l'Hôpital, 75013 Paris, France. Phone: (33) 1 42 17 7401. Fax: (33) 1 42 17 74 11. E-mail: pablogol{at}aol.com.
| |
REFERENCES |
|---|
|
|
|---|
| 1. | Aoki, S., R. Yarchoan, R. V. Thomas, J. M. Pluda, K. Marczyk, S. Broder, and H. Mitsuya. 1990. Quantitative analysis of HIV-1 proviral DNA in peripheral mononuclear cells from patients with AIDS or ARC: decrease of proviral DNA content following treatment with 2',3'-deoxyinosine (ddI). AIDS Res. Hum. Retroviruses 6:1331-1339[Medline]. |
| 2. | Aubin, J. T., D. Boulay, A. Chapus, C. Bréchot, F. Laure, and H. Agut. 1995. Evaluation of a new non-isotopic sandwich hybridization for the detection of HIV1-specific PCR products. Res. Virol. 146:75-79[Medline]. |
| 3. | Bagnarelli, P., S. Menzo, A. Valenza, S. Paolucci, S. Petroni, G. Scalise, R. Sampaolesi, A. Manzin, P. E. Varaldo, and M. Clementi. 1995. Quantitative molecular monitoring of human immunodeficiency virus type 1 activity during therapy with specific antiretroviral compounds. J. Clin. Microbiol. 33:16-23[Abstract]. |
| 4. | Barlow, K. L., J. H. C. Tosswil, J. V. Parry, and J. P. Clewley. 1997. Performance of the Amplicor human immunodeficiency virus type 1 PCR and analysis of specimens with false-negative results. J. Clin. Microbiol. 35:2846-2853[Abstract]. |
| 5. | Bates, D. L. 1987. Enzyme amplification in diagnostics. Trends Biotechnol. 5:204-209. |
| 6. | Binley, J. M., P. J. Klasse, Y. Cao, I. Jones, M. Markowitz, D. D. Ho, and J. P. Moore. 1997. Differential regulation of the antibody responses to Gag and Env proteins of human immunodeficiency virus type 1. J. Virol. 71:2799-2809[Abstract]. |
| 7. | Bobrow, M. N., T. D. Harris, K. J. Shaughnessy, and G. J. Litt. 1989. Catalyzed reporter deposition, a novel method of signal amplification. Application to immunoassays. J. Immunol. Methods 125:279-285[Medline]. |
| 8. | Bollinger, R. C., R. L. Kline, H. L. Franics, M. W. Moss, J. G. Bartlett, and T. C. Quinn. 1992. Acid dissociation increases the sensitivity of p24 antigen detection for the evaluation of antiviral therapy and disease progression in asymptomatic human immunodeficiency virus-infected persons. J. Infect. Dis. 165:913-916[Medline]. |
| 9. | Böni, J., M. Opravil, Z. Tomasik, M. Rothen, L. Bisset, P. Grob, R. Lüthy, and J. Schüpbach. 1997. Simple monitoring of antiretroviral therapy with a signal-amplification-boosted HIV-1 p24 antigen assay with heat-denatured plasma. AIDS 11:F47-F52[Medline]. |
| 10. |
Boom, R.,
C. J. A. Sol,
M. M. M. Salimans,
C. L. Jansen,
P. M. E. Wertheim-van Dillen, and J. van der Noordaa.
1990.
Rapid and simple method for purification of nucleic acids.
J. Clin. Microbiol.
28:495-503 |
| 11. | Bruisten, S. M., P. H. Frissen, P. Van Swieten, P. R. Harrigan, I. Kinghorn, B. Larder, and H. Weigel. 1997. Prospective longitudinal analysis of viral load and surrogate markers in relation to clinical progression in HIV type 1-infected persons. AIDS Res. Hum. Retroviruses 13:327-335[Medline]. |
| 12. | Brun-Vezinet, F., and J. Dormont. 1996. Méthodes et indications. Mesure de la charge virale dans le suivi des patients atteints par le VIH, p. 13-47. In Médecine-sciences, Paris, France. |
| 13. | Caliendo, A. M. 1995. Laboratory methods for quantitating HIV RNA. AIDS Clin. Care 7:89-91. |
| 14. |
Cao, Y.,
F. Valentine,
S. Hojvat, et al.
1987.
Detection of HIV antigen and specific antibodies to HIV core and envelope proteins in sera of patients with HIV infection.
Blood
70:575-578 |
| 15. | Clark, S. J., M. S. Saag, W. D. Decker, S. Campbell-Hill, J. L. Robertson, P. J. Veldkamp, J. C. Kapper, B. H. Hahn, and G. M. Shaw. 1991. High titers of cytopathic virus in plasma of patients with symptomatic primary HIV-1 infection. N. Engl. J. Med. 324:954-960[Abstract]. |
| 16. |
Coombs, R. W.,
D. R. Henrard,
W. F. Mehaffey,
J. Gibson,
E. Eggert,
T. C. Quinn, and J. Phillips.
1993.
Cell-free plasma human immunodeficiency type 1 titer assessed by culture and immunocapture-reverse transcription-polymerase chain reaction.
J. Clin. Microbiol.
31:1980-1986 |
| 17. | Coste, J., B. Montes, J. Reynes, M. Peeters, C. Segarra, J. P. Vendrell, E. Delaporte, and M. Segondy. 1996. Comparative evaluation of three assays for the quantitation of human immunodeficiency virus type-1 RNA in plasma. J. Med. Virol. 50:293-302[Medline]. |
| 18. | Defer, C., H. Agut, A. Garbarg-Chenon, M. Moncany, F. Morinet, F. Vignon, M. Mariotti, and J. J. Lefrère. 1992. Multicentre quality control of polymerase chain reaction for detection of HIV DNA. AIDS 6:659-663[Medline]. |
| 19. | Delamare, C., M. Burgaud, M. J. Mayaux, S. Blanche, A. Doussin, S. Ivanoff, M. L. Chaix, C. Khan, and C. Rouzioux. 1997. HIV-1 RNA detection in plasma for the diagnosis of infection in neonates. The French Pediatric HIV Infection Study Group. J. Acquired Immune Defic. Syndr. Hum. Retrovirol. 15:121-125[Medline]. |
| 20. | Dewar, R. L., et al. 1994. Application of branched DNA signal amplification to monitor human immunodeficiency virus type 1 burden in human plasma. J. Infect. Dis. 170:1172-1179[Medline]. |
| 21. | Ferre, F., A. Marchese, P. C. Duffy, D. E. Lewis, M. R. Wallace, H. J. Beecham, K. G. Burnett, F. C. Jensen, and D. J. Carlo. 1992. Quantitation of HIV viral burden by PCR in HIV seropositive Navy personnel representing Walter Reed stages 1 to 6. AIDS Res. Hum. Retroviruses 8:269-275[Medline]. |
| 22. | Gobbers, E., K. Fransen, T. Oosterlaken, W. Janssens, L. Heyndrickx, T. Ivens, K. Vereecken, R. Schoones, P. van de Wiel, and G. van der Groen. 1997. Reactivity and amplification efficiency of the NASBA HIV-1 RNA amplification system with regard to different HIV-1 subtypes. J. Virol. Methods 66:293-301[Medline]. |
| 23. |
Hammer, S.,
C. Crumpacker,
R. D'Aquila,
B. Jackson,
J. Lathey,
D. Livnat, and P. Reichelderfer.
1993.
Use of virologic assays for detection of human immunodeficiency virus in clinical trials: recommendations of the AIDS Clinical Trials Group Virology Committee.
J. Clin. Microbiol.
31:2557-2564 |
| 24. | Henrard, D. R., S. Wu, J. Phillips, D. Wiesner, and J. Phair. 1995. Detection of p24 antigen with and without immune complex dissociation for longitudinal monitoring of human immunodeficiency virus type 1 infection. J. Clin. Microbiol. 33:72-75[Abstract]. |
| 25. | Iuliano, R., G. Forastieri, M. Brizzi, L. Mecocci, F. Mazzotta, and L. Ceccherini-Nelli. 1997. Correlation between plasma HIV-1 RNA levels and the rate of immunologic decline. J. Acquired Immune Defic. Syndr. Hum. Retrovirol. 14:408-414[Medline]. |
| 26. | Kageyama, S., O. Yamada, and S. S. Mohammad. 1988. An improved method for the detection of HIV antigen in the blood of carriers. J. Virol. Methods 22:125-131[Medline]. |
| 27. |
Kenten, J. H.,
S. Gudibande,
J. Link,
J. J. Willey,
B. Curfman,
E. O. Major, and R. J. Massey.
1992.
Improved electrochemiluminescent laser for DNA probe assays: rapid quantitative assays of HIV-1 polymerase chain reaction products.
Clin. Chem.
38:873-879 |
| 28. | Kestens, L., G. Hoofd, P. L. Gigase, R. Deleys, and G. Van Der Groen. 1991. HIV antigen detection in circulating immune complexes. J. Virol. Methods 31:67-76[Medline]. |
| 29. | Koulinska, I., S. Stakteas, P. Vaz, and J. Barreto. 1997. Diagnosing pediatric AIDS in resource-poor settings: heat-denatured p24 antigen, anti-HIV IgA, or both? J. Acquired Immune Defic. Syndr. Hum. Retrovirol. 16:407-408[Medline]. |
| 30. | Lauré, F., V. Courgnaud, C. Rouzioux, S. Blanche, F. Veber, M. Burgard, C. Jacomet, C. Griselli, and C. Bréchot. 1988. Detection of HIV-1 DNA in infants and children by means of the polymerase chain reaction. Lancet ii:538-541. |
| 31. | Lillo, F. B., M. Mailard, A. Saracco, and O. E. Varnier. 1997. Monitoring antiretroviral activity using ICDp24 and CD4 counts in HIV infection. J. Infect. 35:67-71[Medline]. |
| 32. | MacDonell, C., J. S. Chmiel, L. Poggensee, S. Wu, and J. P. Phair. 1990. Predicting progression to AIDS: combined usefulness of CD4+ lymphocytes counts and p24 antigenemia. Am. J. Med. 89:706-712[Medline]. |
| 33. | Merigan, T. C., G. Skowron, S. A. Bozzette, et al. 1989. Circulating p24 antigen levels and responses to didedoxycytidine in human immunodeficiency virus infection. Ann. Intern. Med. 110:189-194. |
| 34. |
Michael, N. L.,
M. Vahey,
D. S. Burke, and R. R. Redfield.
1992.
Viral DNA and mRNA expression correlate with the stage of human immunodeficiency virus (HIV) type 1 infection in humans: evidence for viral replication in all stages of HIV disease.
J. Virol.
66:310-316 |
| 35. |
Miles, S. A.,
E. Balden,
L. Magpantay,
L. A. Wei,
Leiblein,
D. Hofheinz,
G. Toedter,
E. R. Ehm,
Y. Bryson, and the Southern California Pediatric AIDS Consortium.
1993.
Rapid serologic testing with immune-complex-dissociated HIV p24 antigen for early detection of HIV infection in neonates.
N. Engl. J. Med.
328:297-302 |
| 36. | Moss, A. R., P. Bacchetti, D. Osmond, W. Krampf, R. E. Chaisson, D. Stites, J. Wilber, J. P. Allain, and J. P. Carison. 1988. Seropositivity for HIV-1 and the development of AIDS or AIDS related condition: three year follow-up of the San Francisco General Hospital Cohort. Br. Med. J. 296:745-750. |
| 37. |
Mulder, J.,
N. McKinney,
C. Christopherson,
J. Sninsky,
L. Greenfield, and S. Kwok.
1994.
Rapid and simple PCR assay for quantitation of human immunodeficiency virus type 1 RNA in plasma: application to acute retroviral infection.
J. Clin. Microbiol.
32:292-300 |
| 38. | Nesheim, S., F. Lee, M. L. Kalish, Ocu Cy, M. Swyer, S. Clark, L. Meadows, V. Grimes, R. J. Simonds, and A. Nahmais. 1997. Diagnosis of perinatal human immunodeficiency virus infection by polymerase chain reaction and p24 antigen detection after immune complex dissociation in an urban community hospital. J. Infect. Dis. 175:1333-1336[Medline]. |
| 39. | Nishanian, P., K. R. Huskins, S. Stehn, R. Detels, and J. L. Fahey. 1990. A simple method for improved assay demonstrates that HIV p24 antigen is present as immune complexes in most sera from HIV-infected individuals. J. Infect. Dis. 162:21-28[Medline]. |
| 40. | O'Shea, S., T. Rostrom, A. S. Hamblin, S. J. Palmer, and J. E. Banatvala. 1991. Quantification of HIV: correlation with clinical, virological and immunological status. J. Med. Virol. 35:65-69[Medline]. |
| 41. | Phillips, S. K., B. Parekh, and R. R. George. 1993. Comparison of methods for HIV p24 antigen tests and methods for dissociation of immune complexes, abstr. 174. First National Conference on Human Retroviruses. Washington, D.C . |
| 42. | Saag, M. S. 1997. Use of virologic markers in clinical practice. J. Acquired Immune Defic. Syndr. Hum. Retrovirol. 16(Suppl.1):S3-S13. |
| 43. |
Saag, M. S.,
E. A. Emini,
O. L. Laskin,
J. Douglas,
W. I. Lapidus,
W. A. Schleif,
R. J. Whitley,
C. Hildebrand,
V. W. Byrnes,
J. C. Kappes,
K. W. Anderson,
F. E. Massari,
G. M. Shaw, and the L-697,661 Working Group.
1993.
A short-term evaluation of L-697,661, a non-nucleoside inhibitor of HIV-1 reverse transcriptase.
N. Engl. J. Med.
329:1065-1072 |
| 44. | Schockmel, G. A., S. Yerly, and L. Perrin. 1997. Detection of low HIV-RNA levels in plasma. J. Med. Virol. 51:179-183. |
| 45. | Schüpbach, J., M. Flepp, D. Pontelli, Z. Tomasik, R. Lüthy, and J. Böni. 1992. Heat-mediated immune complex dissociation and enzyme-linked immunosorbent assay signal amplification render p24 antigen detection in plasma as sensitive as HIV-1 RNA detection by polymerase chain reaction. AIDS 10:1085-1090. |
| 46. | Sninsky, J. J., and S. Kwok. 1993. The application of quantitative polymerase chain reaction to therapeutic monitoring. AIDS 7:S29-S34. |
| 47. | Spijkerman, I. J., M. Prins, J. Goudsmit, P. J. Veugelers, R. A. Coutinho, F. Miedema, and F. de Wolf. 1997. Early and late HIV-1 RNA level and its association with other markers and disease progression in long-term AIDS-free homosexual men. AIDS 11:1383-1388[Medline]. |
| 48. | Van Damme, A.-M., J. C. Schmit, and S. Van Dooren. 1996. Quantification of HIV-1 RNA in plasma: comparable results with the NASBA HIV-RNA QT and the Amplicor HIV Monitor Test. J. Acquired Immune Defic. Syndr. Hum. Retrovirol. 13:127-139[Medline]. |
| 49. | Van Gemen, B., T. Kievits, R. Schukkink, D. Van Strijp, L. T. Malek, R. Sooknanan, H. G. Huisman, and P. Lens. 1993. Quantification of HIV-1 RNA in plasma using NASBA during HIV-1 primary infection. J. Virol. Methods 43:177-188[Medline]. |
| 50. | Vella, S. 1995. Immunological and virological markers in HIV infection. AIDS Clin. Care 7:37-40. |
| 51. | Zhang, L. Q., P. Simmonds, C. A. Ludlam, and A. J. Leigh Brown. 1991. Detection, quantification and sequencing of HIV-1 from the plasma of seropositive individuals and from factor VIII concentrates. AIDS 5:675-681[Medline]. |
This article has been cited by other articles:
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| Antimicrob. Agents Chemother. | Clin. Microbiol. Rev. | Infect. Immun. |
|---|---|---|
| J. Clin. Microbiol. | J. Virol. | ALL ASM JOURNALS |