Previous Article | Next Article 
Clinical and Diagnostic Laboratory Immunology, May 2000, p. 336-343, Vol. 7, No. 3
1071-412X/00/$04.00+0
Copyright © 2000, American Society for Microbiology. All rights reserved.
Evaluation of TruCount Absolute-Count Tubes for
Determining CD4 and CD8 Cell Numbers in Human Immunodeficiency
Virus-Positive Adults
Carol T.
Schnizlein-Bick,1,*
John
Spritzler,2
Cynthia L.
Wilkening,2
Janet K. A.
Nicholson,3
Maurice R. G.
O'Gorman,4
Site Investigators,
and
The NIAID DAIDS New Technologies
Evaluation Group
Department of Medicine/Infectious Diseases,
Indiana University School of Medicine, Indianapolis, Indiana
462021; Harvard School of Public Health,
Boston, Massachusetts 021462; Division
of HIV/AIDS, National Center for Infectious Diseases, Atlanta, Georgia
303333; and The Children's Memorial
Hospital and Department of Pediatrics, Northwestern University
Medical School, Chicago, Illinois 606144
Received 26 April 1999/Returned for modification 3 August
1999/Accepted 27 September 1999
 |
ABSTRACT |
A single-platform technology that uses an internal bead standard
and three-color flow cytometry to determine CD4 and CD8 absolute counts
was evaluated for reproducibility and agreement. Values obtained using
TruCount absolute-count tubes were compared to those obtained using a
two-color predicate methodology. Sixty specimens from human
immunodeficiency virus type 1-infected donors were shipped to five
laboratories. Each site also analyzed replicates of 14 human
immunodeficiency virus type 1-infected local specimens at 6 h and
again at 24 h. The interlaboratory variability was significantly
less with TruCount (median difference in percent coefficient of
variation [%CV] between the two methods was
8% and
3% for CD4
and CD8, respectively) than with the predicate method. Intralaboratory
variability was smaller, with a median difference in %CV of
1% for
both CD4 and CD8 with 6-h samples and
2% and
3% for CD4 and CD8,
respectively, with 24-h samples. Use of TruCount for shipped samples
resulted in a median CD4 count change of 7 cells (50th estimated
percentile) when all laboratories and CD4 strata were combined. For
on-site samples, the median CD4 count change was 10 CD4 cells for 6-h
samples and 2 CD4 cells for 24-h samples. Individual site biases
occurred in both directions and cancelled each other when the data were
combined for all laboratories. Thus, the combined data showed a smaller
change in median CD4 count than what may have occurred at an individual
site. In summary, the use of TruCount decreased both the inter- and
intralaboratory variability in determining absolute CD4 and CD8 counts.
 |
INTRODUCTION |
Human immunodeficiency virus type 1 (HIV-1) infects cells that express the CD4 receptor (8) and,
as a result, depletes its host of CD4 lymphocytes (11). This
depletion of CD4 T lymphocytes has been linked to the
immunopathogenesis of HIV infection and progression of the disease
(9, 13). A CD4 count of
200 cells/µl has been included
as an AIDS-defining event (5), as these measurements are
useful predictors for the onset of opportunistic diseases such as
Pneumocystis carinii pneumonia (4). With the
advent of highly active antiretroviral therapy, CD4 T-lymphocyte
measurements have been used to monitor immune reconstitution
(1).
The current predicate methodology for determining absolute CD4
T-lymphocyte counts is dependent upon immunophenotypic identification of cells with fluorescently labeled monoclonal antibodies directed against the CD4 antigen. Relative percentages of CD4 T cells are determined with a flow cytometer. An absolute CD4 count is derived by
multiplying the percentage of lymphocytes that are CD3+
CD4+ by the absolute lymphocyte count determined with a
hematology instrument. However, the overnight shipment of blood may
result in increased intrinsic variability in the absolute lymphocyte count depending on the hematology instrument that is used (10, 16). Therefore, the absolute CD4 count in overnight samples may
have increased variability due solely to the hematological determinants.
The need for precise and reproducible monitoring of CD4 T-lymphocyte
levels in HIV-infected patients has led several companies to develop
simpler methods for measuring absolute CD4 and CD8 T-lymphocyte counts
(2, 7, 14, 15, 17). The new single-platform system developed
by BD Biosciences (San Jose, Calif.) eliminates the need for multiple
technologies (i.e., flow cytometry and hematology) and should be less
expensive than predicate methods when labor, cost and inconvenience of
repeat samples, and hematology costs are considered. TruCount
absolute-count tubes contain a lyophilized pellet that dissolves during
sample preparation, releasing a known number of fluorescent beads. By
gating the bead population during analysis, absolute cell counts can be
readily determined by a simple calculation.
The purpose of this study was to evaluate the single-platform
methodology of TruCount tubes as an alternative method for determining CD4 and CD8 absolute counts and to compare these values with those obtained by predicate methodology. Both reproducibility and agreement with the predicate method were measured on centrally shipped specimens as well as on replicate samples of specimens obtained at individual sites. In addition, the sample stability of on-site replicate 6-h and
24-h paired specimens was investigated.
 |
MATERIALS AND METHODS |
Evaluation sites and instrumentation.
The five participating
laboratories, certified by the National Institute of Allergy and
Infectious Disease (NIAID) flow cytometry proficiency program, were
chosen to represent different geographical locations from within the
United States: two on the east coast, one in the Midwest, and two on
the west coast. All laboratories used a FACScan flow cytometer equipped
with either a Hewlett Packard or Macintosh Quadra computer. For
complete blood count and lymphocyte differential determinations, three
laboratories used Coulter STKS instruments, a fourth laboratory used a
Coulter MD 16 instrument, and the fifth laboratory used a Roche Helios instrument.
Sample collection.
Peripheral blood samples were obtained
only from HIV-1-infected persons. Approval and informed consent were
obtained from all participants. A central contractor, FAST Systems
Inc., Gaithersburg, Md., shipped aliquots of EDTA-anticoagulated blood
from HIV-1-infected donors to individual sites until a total of 60 common specimens were analyzed at all sites for flow cytometry and
hematology. The mailings were received at the sites the day following
the blood drawing and were analyzed on the day of receipt. The
specimens were stratified with regard to CD4 absolute count determined
by the predicate methodology, so that approximately one-third of the
shipped samples were in each stratum: <200, between 200 and 500, and
>500 cells/µl.
In addition, each laboratory obtained EDTA-anticoagulated blood from 14 local HIV-1-infected donors in a single blood drawing consisting of two
7-ml and sixteen 2.5-ml tubes. Eight 2.5-ml EDTA tubes were sent to the
hematology laboratory immediately for replicate determinations. One
7-ml EDTA tube was used for eight replicate determinations of
CD3+ CD4+ and CD3+ CD8+
using the two different flow cytometric methods. The remaining EDTA
tubes were kept overnight at room temperature for 24-h measurements. Donors were prescreened for CD4 count to ensure that seven donors had
CD4 counts of <200 cells/µl and the other seven donors had CD4
counts of
200 cells/µl. Sites continued to process donor specimens
until 14 acceptable 6-h and 24-h paired data sets were obtained.
Antibody staining procedure and data collection.
The
immunophenotyping panels performed on all shipped and on-site specimens
are listed in Fig. 1. All reagents and
computer software used for data acquisition and analysis were supplied by BD Biosciences. For the predicate method, Simultest two-color antibodies were used for staining and SimulSET software was used for
automated data collection. The 1997 Centers for Disease Control revised
guidelines for the performance of CD4 determinations were followed for
samples processed by the two-color, stain, wash, and fix predicate
method (6). For the single-platform method, 20 µl of
TriTEST three-color antibodies and 50 µl of whole blood were added to
bead-containing TruCount tubes. Tubes were incubated for 20 min at room
temperature before 450 µl of FACS Lysing Solution was added. Tubes
were analyzed the same day with CELLQUEST software. The bead population
and the CD45 lymphocyte versus side scatter population were manually
gated. The absolute count using TruCount tubes was calculated from the
appropriate dot plot values entered into a spreadsheet that was
formatted to use the formula [(no. of events in quadrant containing
cell population)/(no. of events in absolute-count bead region [R2])] × [(total no. of absolute-count beads)/(test volume [50 µl])].

View larger version (16K):
[in this window]
[in a new window]
|
FIG. 1.
Antibody staining profiles. FITC, fluorescein
isothiocyanate; PE, phycoerythrin; PerCP, peridinin chlorophyll
protein; MsIgG, mouse immunoglobulin G.
|
|
Hematology measurements.
The five hematology laboratories
that participated in this study maintained performance that conformed
to accepted standards of practice (e.g., College of American
Pathologists and National Committee for Clinical Laboratory Standards).
All hematology samples were drawn at the same time as the specimens for
flow cytometry. For shipped specimens, the hematology measurements were
performed within 33 h of specimen draw. For on-site specimens, the
measurements were performed within 7 h of draw for same-day
specimens and within 33 h of draw for the paired 24-h specimen
data. White blood cell (WBC) and leukocyte differential (including
percent lymphocyte) counts were performed on an automated instrument.
If the specimen was rejected or flagged in the lymph region by the
machine, the value was flagged in the database spreadsheet and
subsequently eliminated from the study analyses. A cell designated as
an atypical lymph or large unstained cell was included in the total
lymphocyte number.
Analyses.
Criteria for accepting data obtained with SimulSET
software included the following: (i) gated lymphocyte purity >85%,
(ii) lymphocyte recovery within the gate >90%, and (iii) differences in the CD3 percentages between the CD3+ CD4+
and CD3+ CD8+ tubes
7%. Data from individual
sites were entered into a spreadsheet designed by the Statistical Data
Analysis Center, Harvard School of Public Health, Boston, Mass., and
imported into a central database for analyses. Comparisons of the
variability of the TruCount method versus the predicate method were
based on the Wilcoxon signed-rank test (12) applied to the
differences (TruCount method minus predicate method) in the percent
coefficient of variation (%CV) of reported CD4 and CD8 counts for each
specimen (between-laboratory %CV in the case of the centrally shipped
specimens and within-replicate %CV in the case of local donor
specimens). The accuracy of the CD4 and CD8 counts determined by the
TruCount method versus the predicate method was tested by the Wilcoxon
signed-rank test applied to the differences (TruCount method minus
predicate method) in reported CD4 and CD8 counts for each centrally
shipped specimen at each laboratory. In the case of CD4 and CD8 counts
on replicates from local donors, a log rank test stratified by donor
was used. For the primary endpoints, statistical significance is
defined as P < 0.05. The primary endpoints for both
CD4 and CD8 counts, and combining all CD4 strata, are (i)
intralaboratory variability in all laboratories combined using local
specimens 24 h old, (ii) interlaboratory variability using
centrally shipped specimens, (iii) intralaboratory agreement using
local specimens 24 h old, and (iv) intralaboratory agreement using
centrally shipped specimens. The P values for the secondary
endpoints are exploratory only. Tertiary analyses on the CD4 and CD8
subset percentages were carried out similarly.
 |
RESULTS |
Interlaboratory variability for shipped specimens.
CD4 and CD8
absolute counts were obtained for 60 samples shipped to five different
laboratories. Statistical analyses of the %CVs of the TruCount method
minus the %CVs of the predicate method were performed on the database
as a whole, on the different CD4 strata, and on individual site data.
Table 1 shows that the median %CV for
the TruCount method was 8% and 3% less than the predicate method for
CD4 and CD8, respectively. When analyzed with regard to CD4 stratum,
the <200 cells/µl group showed a significant 23% (CD4) and 9%
(CD8) median difference in %CVs for the two methods. These large
differences were due to the large median %CVs for the predicate method
for samples with a CD4 count of <200 cells/µl. For the remaining two
CD4 strata, the median differences between methods were again
significantly lower (3 to 7%), favoring the TruCount method.
Intralaboratory variability for 6-h and 24-h replicate
samples.
Each of the five sites solicited 14 donors whose samples
were analyzed as eight replicates at 6 h and eight replicates at 24 h. Therefore, a total of 70 paired samples (35 with a CD4 count of <200 cells/µl and 35 with a CD4 count of
200 cells/µl)
constituted the database. Table 2 shows
that the median difference in %CVs for 6-h replicate samples was
1%
(%CV for TruCount
%CV for predicate method) for both CD4 and
CD8 counts by both methods. When analyzed with regard to CD4 strata,
specimens with CD4 counts of <200 cells/µl showed significant
differences in the %CVs for CD4 and CD8 counts (
3 and
1%,
respectively). For samples with
200 CD4 cells/µl, no differences in
median %CVs between the two methods were seen. Individual site
performance reflected what was observed for the database as a whole.
For the on-site replicate samples held overnight, the TruCount tubes
generated overall median %CV differences that were less than the
predicate method values for both CD4 and CD8 counts (2 and 3%,
respectively) (Table 3). When the samples
with a CD4 count of <200 cells/µl were analyzed, the median
differences in %CVs were 3 and 4%, TruCount values being less than
predicate method values for CD4 and CD8 counts, respectively. Samples
with a CD4 count of
200 cells/µl showed significant median
differences with %CVs of
2% for CD4 counts but an insignificant
1% for CD8 counts. In general, individual site performance reflected
what was observed for the database as a whole.
Agreement for CD4 and CD8 absolute counts with the predicate method
for shipped samples.
Agreement between methods was assessed by
subtracting the absolute count obtained by the predicate method from
the absolute count obtained by the TruCount method for the same sample,
and the 10th, 50th, and 90th percentiles of the differences were
estimated. For shipped samples, with all CD4 strata and laboratories
combined, the median of the differences for CD4 and CD8 counts were 7 and
51 cells, respectively (Table 4).
With all CD4 strata combined, the median of the differences for CD4
counts from individual laboratories ranged from
34 to 25 cells.
Individual site biases were noted for both CD4 and CD8 counts between
the two methods. Laboratories A and B obtained lower values for CD4
counts using TruCount, while laboratories C, D, and E obtained higher
values with the tubes.
Agreement of CD4 and CD8 absolute counts for on-site replicate
samples.
For 6-h replicate samples, the median of the differences
for CD4 and CD8 counts between TruCount and the predicate method for
all CD4 strata and laboratories were 10 and 23 cells, respectively (Table 5). The median of the differences
for CD4 and CD8 counts for the 24-h replicate samples were 2 and
21
cells, respectively; these differences were similar in direction and
magnitude to those obtained for the shipped samples (CD4 results
similar to and CD8 TruCount results lower than those with the predicate
method). Individual site biases were noted for both CD4 and CD8 counts between the two methods for both the 6-h and the 24-h replicate samples. Laboratory B obtained lower values for CD4 counts using TruCount, while laboratories A, C, D, and E obtained higher values with
this method.
Interlaboratory variability of subset percentages.
In an
attempt to estimate the contribution of the intrinsic variability of
the WBC and leukocyte differential counts to the combined CD4
absolute-count variability, the subset percentages (CD3+
CD4+ and CD3+ CD8+) obtained by
both methods were evaluated. For both shipped and 24-h on-site
replicate samples, the median %CVs obtained for CD4 and CD8 subset
percentages were significantly less using the TruCount method than the
predicate method (Table 6). The median
difference in %CVs for CD4 subset percentage was almost half for
shipped samples and about one third less for 24-h on-site replicate
samples using the TruCount tubes. For CD8 subset percentages, these
differences were of a lower magnitude: the median difference in %CVs
was about 1% lower using the TruCount tubes for both shipped and 24-h
on-site replicate samples. Similar differences were seen when the data were analyzed with regard to individual CD4 count stratum.
Sample stability of paired specimens analyzed at 6 and 24 h.
An indicator of sample stability was determined by subtracting
the 6-h count from the 24-h count for the same paired samples analyzed
by the same method. Therefore, differences in sample stability were
determined separately for the predicate and TruCount methods. For CD4
counts, the median of the difference in fresh and day-old samples using
TruCount tubes was
2 cells, compared with 4 cells using the predicate
method when all laboratories and CD4 strata were combined (Table
7). In other words, TruCount tubes gave
slightly higher CD4 values at 6 h than at 24 h, while the
predicate method gave slightly lower values at 6 h than at 24 h. Similar differences were observed when the CD4 strata were evaluated. Each laboratory had its individual bias for paired samples
at 6 and 24 h. Table 8 shows a
similar data profile for 6-h and 24-h paired CD8 counts. When all
strata and laboratories were combined, the median of the differences in
CD8 counts using TruCount was higher for fresh specimens (15 cells) but
lower for day-old specimens (35 cells). Similar individual site biases
were observed for CD8 counts for the paired samples at both time
points. In general, the stability data showed that use of TruCount
tubes resulted in higher CD4 and CD8 counts at 6 h than at 24 h.
 |
DISCUSSION |
The current method for the measurement of CD4 absolute counts is
expensive and time-consuming and requires multiple manipulations. In
addition, most laboratories require separate tubes of blood for the
flow cytometry and hematology measurements. Since the hematology
measurement is often sensitive to small changes in the blood
components, the intrinsic variability of this measurement is especially
difficult to minimize. Since the interlaboratory variability in CD4
absolute count for shipped specimens has been unacceptably large, the
values for overnight samples may not be reliable. Another confounding
factor is that different hematology instruments may have biases toward
higher or lower lymphocyte counts (3, 18). This poses a
serious problem for patients on HIV-1 intervention protocols if they
change where their laboratory CD4 determinations are made. The high
variability between any two laboratories may make longitudinal
comparisons of CD4 counts inaccurate. The availability of
single-platform technologies, which determine CD4 or CD8 absolute
counts using only flow cytometry, would decrease this problem and make
absolute counts between institutions less variable.
In the current study, which compared single-platform TruCount tubes
with a multiplatform predicate method, both the interlaboratory and
intralaboratory variability was significantly less using TruCount tubes
for both shipped and on-site replicate samples. At the time that the
schema was developed, the predicate method in general use did not use
CD45 gating. The purpose of this study was to determine the differences
in reproducibility and agreement that would occur if a laboratory
switched from their current two-color predicate method to a
single-platform method. For shipped samples, the variability between
CD4 absolute counts using TruCount tubes was about half that using the
predicate method. One could argue that this decrease in variability was
largely due to the CD45 gating strategy used in the TruCount method and
that this decrease in variability could also be achieved with any
multiplatform method that used three-color and CD45 gating. If this
were true, the difference in variability observed in the subset
percentage data for the two methods should account for the majority of
the decreased variability in the absolute-count data. However, an
analysis of the CD4 subset percentages showed that only a small portion
of the decreased variability could be attributed to the use of CD45 gating. This suggests that the TruCount method, in addition to improving the precision of determining lymphocyte subsets over the
predicate method, also eliminated the intrinsic variability contributed
by the hematology measurements.
When the difference in absolute counts for the same sample by the two
methods was calculated, small differences were detected for the
database as a whole. For example, the agreement for the 60 shipped
samples was a median CD4 count change of 7 cells. Likewise, for on-site
replicate samples, the agreement was 10 CD4 cells for 6-h samples and 2 cells for 24-h replicate samples. However, these values were misleading
because individual sites had biases in the absolute counts that varied
in both magnitude and direction. Individual site evaluations showed
significant changes in absolute-count values between the two methods,
but since site subset percentages did not show these directional biases
(data not shown), one must conclude that the site's bias in the
absolute-count determinations arose from the site's hematology
instrument. For example, laboratory B used a Roche Helios hematology
instrument, and that site's predicate-method absolute counts were
consistently higher than those obtained by the TruCount method
determinations compared with the other sites, which used Coulter
hematology instruments. Thus, it is important that a site perform a
comparative study to determine whether a bias in their reported CD4 and
CD8 absolute counts will occur if they switch to a single-platform methodology.
Single-platform technology can be more cost-effective than the
predicate method when the cost of the WBC and lymphocyte differential counts, the time saved with reduced sample manipulation, and the cost
and inconvenience of redrawing blood are considered. For example, the
1999 BD Biosciences list price for the antibody reagents used in the
predicate method Simultest panel cost $32.30 per patient test. In
comparison, the TriTEST antibody reagents cost $25.40 per patient
test. TruCount tubes add $9.60 to the cost of the TriTEST reagents, to
bring the single-platform cost to $35.00 for a patient test.
Considering that a WBC and lymphocyte differential determination costs
more than $2.70 (TruCount minus Simultest costs) and often more than
$9.60 (TruCount minus TriTEST costs), the single-platform determination
is more cost-effective than the multiplatform methods for reagents
alone. In addition, many flow cytometry laboratories are not able to
receive hematology laboratory reports by direct data transmission.
Since the hematology results must be received by the flow cytometry
laboratory before CD4 and CD8 absolute counts can be calculated,
misplaced reports can result in absolute-count reporting delays.
Occasionally, the tube of blood for the hematology laboratory is not
drawn and the CD4 and CD8 absolute counts cannot be reported for the
patient sample. In other circumstances, sample quality is so poor that the hematology instrument cannot perform a reliable lymphocyte differential, and again absolute counts cannot be calculated. When
hematology values are unattainable for these various reasons, bead-based, single-platform technology can provide meaningful absolute-count data.
Some single-platform methods do not directly determine lymphocyte
percentages and only give CD4 and CD8 absolute counts. Examples include
the FACSCount system, the Ortho Cytoron Absolute system, the Zymmune
CD4/CD8 cell monitoring kit from Zynaxis, Inc. (7), volumetric capillary cytometry from Biometric Imagining, Inc. (15), and the TRAx CD4 test kit from T Cell Diagnostics,
Inc. (17). However, TruCount tubes and Flow-Count
fluorospheres (Beckman Coulter) provide lymphocyte subset percentages
as well as absolute counts. The determination of both of these clinical
parameters is important because the lymphocyte subset values are often
required for monitoring pediatric HIV-positive populations.
In the present study, the predicate method used forward light scatter
versus side scatter in comparison to the TruCount method, which used
CD45 lymphocyte gating for sample acquisition and analysis. CD45
lymphocyte gating is also routinely used today by flow cytometric laboratories performing three-color, multiplatform analyses and is not
unique to the bead-based, single-platform technology used in this
study. Degenerative samples that must be analyzed with a gating
strategy based on forward light scatter versus side scatter often fail
to meet acceptable gating criteria and result in blood specimens having
to be retested or redrawn. For example, in order for the 60 shipped
samples to be analyzable at all five sites, a total of 90 specimens
were actually sent to the laboratories. The majority of extra specimens
were needed because of unacceptable gating criteria for the shipped
samples by the predicate method. However, for those samples that could
not be analyzed with the predicate method, CD4 and CD8 absolute counts
were almost always obtained by using CD45 lymphocyte gating. The
TruCount method had an additional advantage over a three-color
multiplatform method in that it could "rescue" samples for which an
absolute count could not be calculated because of invalid or incomplete
hematology values.
In summary, the TruCount method gave better reproducibility and
agreement for both shipped and on-site replicate samples than a
multiplatform predicate method. In addition to being more
cost-efficient, valid absolute cell counts for degenerative samples
could be consistently obtained by the single-platform method.
Therefore, the results of this multisite study support the use of
bead-based, single-platform technology for routine clinical assessment
of CD4 and CD8 absolute cell counts.
 |
ACKNOWLEDGMENTS |
This work was funded in part by the NIAID Immunophenotyping
Quality Assessment Program, contract NO-AI-45175.
We thank BD Biosciences (Ann Shiba and Jim Lowder) for additional
financial support and contribution of reagents and training of site
personnel. Special appreciation is also extended to the site
technologists Eileen Bessent, Gina Bonifacio, Todd Christian, David
Devernoe, Fred Menendez, and Kristen Stank, who prepared and analyzed
the samples.
 |
FOOTNOTES |
*
Corresponding author. Mailing address: Department of
Medicine/Infectious Diseases, Indiana University School of Medicine, 1001 W. 10th St. Room OPW 430, Indianapolis, IN 46202-2879. Phone: (317) 630-6971. Fax: (317) 630-7522. E-mail:
cschnizl{at}iupui.edu.
S. Mudzinski, Albany Medical College; S. Peters, Georgetown
University Hospital; S. Plaeger, University of California, Los Angeles,
School of Medicine; C. Schnizlein-Bick, Indiana University School of
Medicine; C. Spina, University of California, San Diego, School of
Medicine; M. Waxdal and C. Monical, FAST Systems; J. Lowder and A. Shiba, BD Biosciences.
E. Bessent, University of California, San Diego; A. Donnenberg and
S. Douglas, Children's Hospital of Philadelphia; F. Mandy, Bureau for
HIV/AIDS and STD, LCDC, Health Canada; J. Nicholson, Centers for
Disease Control and Prevention; M. O'Gorman, Northwestern University
Medical School; S. Plaeger, University of California, Los Angeles; K. Reimann, J. Spritzler, and C. Wilkening, Harvard School of Public
Health; J. Schmitz, University of North Carolina; C. Schnizlein-Bick,
Indiana University; J. Kagan and D. Livnat, Division of AIDS, National
Institute of Allergy and Infectious Diseases.
 |
REFERENCES |
| 1.
|
Autran, B.,
G. Carcelain,
T. S. Li,
C. Blanc,
D. Mathez,
R. Tubiana,
C. Katlama,
P. Debre, and J. Leibowitch.
1997.
Positive effects of combined antiretroviral therapy on CD4+ T cell homeostasis and function in advanced HIV disease.
Science
277:112-116[Abstract/Free Full Text].
|
| 2.
|
Bene, M. C.,
M. N. Kolopp Sarda,
J. El Kaissouni,
A. De March Kennel,
C. Mole,
C. Kohler, and G. C. Faure.
1998.
Automated cell count in flow cytometry: a valuable tool to assess CD4 absolute levels in peripheral blood.
Am. J. Clin. Pathol.
110:321-326[Medline].
|
| 3.
|
Bentley, S. A.,
A. A. Johnson, and C. A. Bishop.
1993.
A parallel evaluation of four automated hematology analyzers.
Am. J. Clin. Pathol.
100:626-632[Medline].
|
| 4.
|
Centers for Disease Control.
1992.
Recommendations for prophylaxis against Pneumocystic carinii pneumonia for adults and adolescents infected with human immunodeficiency virus.
Morbid. Mortal. Weekly Rep.
41(RR-4):1-11.
|
| 5.
|
Centers for Disease Control.
1992.
Revised classification system for HIV infection and expanded surveillance case definition for AIDS among adolescents and adults.
Morbid. Mortal. Weekly Rep.
41(RR-17):1-35.
|
| 6.
|
Centers for Disease Control.
1997.
1997 revised guidelines for performing CD4+ T-cell determinations in persons infected with human immunodeficiency virus (HIV).
Morbid. Mortal. Weekly Rep.
46(RR-2):1-29[Medline].
|
| 7.
|
Denny, T. N.,
B. D. Jensen,
E. I. Gavin,
A. G. Louzao,
F. A. Vella,
J. M. Oleske, and W. Wong.
1995.
Determination of CD4 and CD8 lymphocyte subsets by a new alternative fluorescence immunoassay.
Clin. Diagn. Lab. Immunol.
2:330-336[Abstract].
|
| 8.
|
De Wolf, F.,
M. Roos,
J. M. A. Lange,
J. T. M. Houweling,
R. A. Coutinho,
J. van der Noordaa,
P. T. Schellekens, and J. Goudsmit.
1988.
Decline in CD4+ cell numbers reflects increase in HIV-1 replication.
AIDS Res. Hum. Retroviruses
4:433-440[Medline].
|
| 9.
|
Fahey, J. L.,
J. M. G. Taylor,
R. Detels,
B. Hoffman,
R. Melmed,
P. Nishanian, and J. V. Giorgi.
1990.
The prognostic value of cellular and serological markers in infection with human immunodeficiency virus type 1.
N. Engl. J. Med.
322:166-172[Abstract].
|
| 10.
|
Koepke, J. A., and A. L. Landay.
1989.
Precision and accuracy of absolute lymphocyte counts.
Clin. Immunol. Immunopathol.
52:19-27[CrossRef][Medline].
|
| 11.
|
Lang, W.,
H. Perkins,
R. E. Anderson,
R. Royce,
N. Jewell, and W. Winkelstein, Jr.
1989.
Patterns of T-lymphocyte changes with human immunodeficiency virus infection: from seroconversion to the development of AIDS.
J. Acquir. Immune Defic. Syndr.
2:63-69.
|
| 12.
|
Lehman, E. L., and H. J. M. D'Abrera.
1975.
Non-parametrics: statistical methods based on ranks.
Holden Day, Inc., Oakland, Calif.
|
| 13.
|
Masur, H.,
F. P. Ognibene,
R. Yarchoan,
J. H. Shelhamer,
B. F. Baird,
W. Travis,
A. F. Suffredini,
L. Deyton,
J. A. Kovacs,
J. Falloon,
R. Davey,
M. Polis,
J. Metcalf,
M. Baseler,
R. Wesley,
V. J. Gill,
A. S. Fauci, and H. C. Lane.
1989.
CD4 counts as predictors of opportunistic pneumonias in human immunodeficiency virus (HIV) infection.
Ann. Intern. Med.
111:223-231.
|
| 14.
|
Nicholson, J. K. A.,
W. M. Velleca,
S. Jubert,
T. A. Green, and L. Bryan.
1994.
Evaluation of alternative CD4 technologies for the enumeration of CD4 lymphocytes.
J. Immunol. Methods
177:43-54[CrossRef][Medline].
|
| 15.
|
O'Gorman, M. R. G.,
R. Gelman Site Investigators, et al.
1997.
Inter- and intrainstitutional evaluation of automated volumetric capillary cytometry for the quantitation of CD4- and CD8-positive T lymphocytes in the peripheral blood of persons infected with human immunodeficiency virus.
Clin. Diagn. Lab. Immunol.
4:173-179[Abstract].
|
| 16.
|
Paxton, H., and T. Bendele.
1993.
Effect of time, temperature, and anticoagulant on flow cytometry and hematological values.
Ann. N. Y. Acad. Sci.
677:440-443[Medline].
|
| 17.
|
Paxton, H.,
M. Pins,
G. Denton,
A. D. McGonigle,
R. S. Meisner, and J. P. Phair.
1995.
Comparison of CD4 cell count by a simple enzyme-linked immunosorbent assay using a TRAx CD4 test kit and by flow cytometry and hematology.
Clin. Diagn. Lab. Immunol.
2:104-114[Abstract].
|
| 18.
|
Robinson, G.,
L. Morgan,
M. Evans,
S. McDermott,
S. Pereira,
M. Wansbrough-Jones, and G. Griffin.
1992.
Effect of type of haematology analyser on CD4 count.
Lancet
340:485[CrossRef][Medline].
|
Clinical and Diagnostic Laboratory Immunology, May 2000, p. 336-343, Vol. 7, No. 3
1071-412X/00/$04.00+0
Copyright © 2000, American Society for Microbiology. All rights reserved.