Clinical and Diagnostic Laboratory Immunology, March 2002, p. 329-332, Vol. 9, No. 2
1071-412X/02/$04.00+0 DOI: 10.1128/CDLI.9.2.329-332.2002
Copyright © 2002, American Society for Microbiology. All Rights Reserved.
Alidex, Inc., Redmond, Washington,1 Research Oriented Biostatistics, Inc., Mundelein, Illinois,2 Focus Technologies, Cypress, California3
Received 10 October 2001/ Returned for modification 20 December 2001/ Accepted 26 December 2001
|
|
|---|
|
|
|---|
The PolyTiter Immunofluorescent Titration system (PolyTiter system; Polymedco, Inc.) was developed to facilitate ANA-IFA testing by providing an endpoint titer directly from a 1:40 dilution. Based on the principle that the intensity of the fluorescence on the ANA slide is related to the dilution endpoint of the specimen, the PolyTiter allows the technologist to regulate the light source in a linear manner so that the intensity of specimen staining can be related to the intensity of established calibrators. By controlling the illumination of an IFA slide in a manner parallel to specimen serial dilution, the system supports establishment of the titer from a single dilution. Unlike automated systems, in which all available light is measured, the technologist decides which features on the slide are to be evaluated by looking at the selected object (cytoplasm, nucleoli, etc). The ability of the user to change focus or objectives is unaffected, and the function of the microscope is unchanged regardless of whether the system is turned on or off. Therefore, the system does not require a dedicated fluorescent microscope. As with conventional ANA testing, the PolyTiter system is technician dependent, allowing the detection and characterization of specimen patterns.
This study was conducted to assess the performance characteristics of the PolyTiter system.
|
|
|---|
![]() View larger version (48K): [in a new window] |
FIG. 1. PolyTiter system hardware.
|
A calibrator curve is plotted from the assigned calibrator concentrations (y axis) as a function of their corresponding PolyTiter values (x axis). The curve is either drawn point to point by the software or plotted by hand. The PolyTiter value for a given patient sample or kit control is located on the plot, and the corresponding titer is read from the graph. Specimens with titers greater than 2,560 may be reported as >2,560 or may be further diluted prior to reassaying. In the case of specimens with mixed patterns, a PolyTiter value for each individual pattern may be determined and the results reported.
The PolyTiter software application was created as an adjunct to the PolyTiter system to automate the generation of the calibration curve, determining control and specimen endpoint dilutions, and providing the results in a printable report format. PolyTiter values may be accepted into the computer software immediately from the control pad or recorded by the operator for input and data reduction after all the sample wells have been evaluated. The PolyTiter system is fully operational without the use of the software.
Reagents and equipment. DiaSorin Anafluor indirect fluorescent antibody tests were used for all ANA assays. Kallestad Laboratories Quantafluor fluorescent tests were included in the analysis comparing two manufacturer's HEp-2 slides. ANA kits were used according to the appropriate manufacturer's instructions. Study IF microscopes included an Olympus BH-Z, a Leitz Diaplan, and an Olympus BX 40F-3.
Analyses. Precision testing was conducted using a specimen set composed of three randomly ordered blind-coded specimens representing low (1:80), medium (1:320), and high (1:1,280) ANA titers. Testing was performed in one run per day by each of two technologists who used the PolyTiter to evaluate 10 replicates of the three specimens on five sequential days, a total of 50 replicates per specimen per reader. Positive and negative controls were included on each of the five slides each day, for a total of 25 replicates per reader. One reader performed all testing with one lot of ANA calibrators; the other reader included calibrators from two additional lots each day and read the result of testing for each replicate from three separate calibrator curves.
The average of three PolyTiter values per well was used to determine the titer from which overall agreement (defined as ±2 dilutions) was made against the reference result. Percent agreement was calculated for each set of replicate specimen and control values. In addition, a nested-components-of-variance approach was used to evaluate reader-to-reader, between-run (day), and within-run (day) precision using the average PolyTiter unit value and the first PolyTiter value only (single read). To evaluate lot-to-lot precision, the pooled estimate of the between-day variation across lots for percent agreement, defined as (SD2Lot 1 + SD2Lot 2 + SD2Lot 3)/3 was calculated, and the lot-to-lot between-run (day) precision was determined for each sample using a coefficient of variation (CV) calculation, defined as [(pooled estimate of between-run variation)1/2/(mean % agreement across lots)] x 100 or [(individual lot between-run variation)1/2/(mean % agreement within an individual lot)] x 100.
A comparison between the PolyTiter and conventional serial dilution in the determination of endpoint titer was conducted at three laboratories, located in California, New Jersey, and Minnesota, using identical 125-member serum specimen sets. Specimens were identified by a commercial source as 90 specimens positive for ANA and 35 negative specimens. Positive specimens included single and multiple patterns, with some rare or unusual patterns as well as some cytoplasmic staining (Table 1). The specimen sets for each site were further divided, with each subset randomly ordered, blind-coded, and designated to be tested by the PolyTiter or by conventional serial dilution.
|
View this table: [in a new window] |
TABLE 1. Specimen set (n = 125) identified by the commercial source for comparison of PolyTiter and serial dilution for determination of endpoint titer
|
1,280 for these specimens. All specimens and controls tested with the PolyTiter system were diluted 1:40. Three PolyTiter readings were performed in each well using the five-step-mode function; thus, PolyTiter readings could range from 0 to 100 in units of 5. Endpoint titer reports of negative, 40, 80, 160, 320, 640, 1,280, 2,560, and >2,560 were possible with the PolyTiter system. Endpoint agreement between the two methods was defined as ±2 dilutions. To evaluate the method comparison, the exact binomial probabilities derived from a one-sample binomial test and corresponding 95% confidence intervals (CI) were computed for the positive and negative samples by site. The null hypothesis tested for each of these comparisons (i.e., positive and negative titer samples) was H0: p = p0 versus H1: p < p0, where p is the observed acceptance proportion and p0 is the expected acceptance proportion (i.e., 95% for positive titer samples and 99% for negative titer samples). Statistical significance was established as P < 0.05.
The 90 specimens identified as positive by the commercial source were also tested at one of the sites with a second manufacturer's ANA kit, and the results of each IFA were used to determine the observed proportions for both assays. A one-sample test for binomial proportions using a one-sided alternative was used for the statistical comparisons. The null hypothesis tested for each of the statistical comparisons was H0: p = p0 versus H1: p < p0, where p is the observed acceptance proportion and p0 is the expected acceptance proportion (i.e., 95% - 10% = 85% for positive titer samples).
|
|
|---|
The results of the precision evaluation using PolyTiter units are provided in Table 2, where the nested-components-of-variance approach yielded a CV of less than 10% for all sources of variation. As expected, the CVs were slightly higher (1 to 2.5%) for a single PolyTiter reading versus the average; however, results remained within acceptable limits, with neither method consistently higher or lower than the other. Furthermore, the variance component in the lot-to-lot analysis was equal to zero for all samples, indicating that all of the variation exhibited was due to run-to-run differences.
|
View this table: [in a new window] |
TABLE 2. Precision testing of PolyTiter unit values
|
|
View this table: [in a new window] |
TABLE 3. Endpoint titer agreement between the PolyTiter and serial dilution methods for 125 samples
|
|
View this table: [in a new window] |
TABLE 4. Titer agreement between conventional serial dilution and the PolyTiter system for 125 samples
|
|
View this table: [in a new window] |
TABLE 5. Pattern agreement between the PolyTiter system and serial dilution for multiple patterns as identified by site
|
|
|
|---|
The PolyTiter system eliminates many of the drawbacks of conventional ANA-IFA while preserving its familiar report format. Regulation of the light source to give barely visible fluorescence allows a more objective determination of endpoint titer. Training of new technologists to determine endpoint titers consistently and accurately is more easily accomplished for the PolyTiter system compared to the traditional serial dilution approach. While the identification of staining patterns and the determination of endpoint titers of mixed pattern staining are not affected, the reader must still be properly trained in pattern recognition.
The PolyTiter system demonstrates acceptable precision and compares positively to conventional serial dilution in the determination of endpoint dilutions while using the same equipment and reagents currently used in the clinical laboratory. The ability of the PolyTiter system to supply an endpoint titer from a single serum dilution offers major advantages over conventional ANA titration methods. In addition to the obvious and significant reductions in reagent costs, those laboratories that perform a single dilution screen followed by titration of positives should see improvements in turnaround time using the PolyTiter system. Furthermore, use of the PolyTiter system should result in reduced technologist labor and fatigue.
Additional studies are needed to analyze specific cost reductions and to determine if the PolyTiter system may also offer increased intra- and/or interlaboratory reproducibility.
|
|
|---|
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||