We evaluated the Ortho TRIO-Cytoronabsolute system for determining
absolute CD4 counts. The CD4 counts in our blood specimens from 100 individuals ranged from 3 to 1,962; the percent CD4 ranged from 1.3 to
62.2, respectively. The TRIO system was biased toward lower absolute
counts than a combination of flow cytometry and hematology but showed
no bias in percent CD4 calculations.
1071-412X/98/$04.00+0
Copyright © 1998, American Society for Microbiology. All rights reserved.
Absolute CD4 Counts Obtained by a Three-Color Flow-Cytometric
Method without the Use of a Hematology Analyzer
*