Previous Article | Next Article ![]()
Clinical and Diagnostic Laboratory Immunology, November 2001, p. 1145-1149, Vol. 8, No. 6
Division of Applied Research and
Technology1 and Division of Surveillance
Hazard Evaluations and Field Studies,2 National
Institute for Occupational Safety and Health, Centers for Disease
Control and Prevention, Public Health Service, Department of Health and
Human Services, Cincinnati, Ohio 45226, and Division of
Allergy and Clinical Immunology, Johns Hopkins Asthma and Allergy
Center, Johns Hopkins University School of Medicine, Baltimore,
Maryland 212243
Received 21 May 2001/Returned for modification 7 August
2001/Accepted 22 August 2001
Receiver operating characteristics (ROC) analyses to evaluate and
compare the diagnostic accuracy of Food and Drug Administration (510K)-cleared natural rubber latex (NRL)-specific immunoglobulin E
(IgE) antibody immunoassays have not been performed using
well-characterized skin-testing reagents. Sera were collected from 311 subjects (131 latex puncture skin test [PST] positive and 180 PST
negative). All masked, coded sera were analyzed for latex-specific IgE
antibodies in the Diagnostic Products Corporation microplate AlaSTAT,
HYCOR HY-TEC RAST, and Pharmacia-Upjohn CAP System RAST FEIA
(CAP). Diagnostic accuracy was evaluated using GraphRoc for Windows
software to construct and analyze ROC curves in relation to the
subjects' PST status and the results of the immunoassays. The ROC
areas under the curve (AUCs) ± standard error based on PST for
the three diagnostic tests were 0.858 ± 0.024, 0.869 ± 0.024, and 0.924 ± 0.017, respectively, for AlaSTAT, CAP, and
HY-TEC. The HY-TEC system had a significantly greater AUC based on PST
than those observed for AlaSTAT (P < 0.05) and CAP
(P < 0.05) analyses. When the diagnostic tests
were probed as to the cutoffs giving maximal diagnostic efficiency
compared to PST, CAP and AlaSTAT yielded values of <0.35 kU of
allergen IgE (kUA)/liter and <0.35 kU/liter while
the HY-TEC assay yielded 0.11 kU/liter. The diagnostic efficiencies based on PST in our cohort at these cutoffs were 87.1, 88.1, and 88.7%, respectively. The HY-TEC assay had a significantly greater AUC
than CAP and AlaSTAT using PST as a diagnostic discriminator in our
cohort. When the HY-TEC system was probed at its maximally efficient
cutoff (0.11 kU/liter) versus HYCOR's recommended cutoff of 0.05 kU/liter, a loss of sensitivity of 8.4% was observed with a gain in
specificity of 19.5%.
1071-412X/01/$04.00+0 DOI: 10.1128/CDLI.8.6.1145-1149.2001
Receiver Operating Characteristics Analyses of Food
and Drug Administration-Cleared Serological Assays for Natural Rubber
Latex-Specific Immunoglobulin E Antibody
*
Corresponding author. Mailing address: Division of
Applied Research and Technology, National Institute for Occupational
Safety and Health, MS-C26, 4676 Columbia Parkway, Cincinnati, OH 45226. Phone: (513) 533-8196. Fax: (513) 533-8494. E-mail:
reb4{at}cdc.gov.
This article has been cited by other articles:
Copyright © 2009 by the American Society for Microbiology. For an alternate route to Journals.ASM.org, visit: http://intl-journals.asm.org | More Info»